Nocturnal Blood Pressure Dipping in Adults: A Comparison Between Cuffless and Ambulatory Blood Pressure Monitoring.
C Hove, T M Seeberg, K G Bøtker-Rasmussen, E Westgaard, S Heimark, B Waldum-Grevbo, A C K Larstorp
Department of Nephrology, Oslo University Hospital, Ullevål, Oslo, Norway. · Aidee Health AS, Bærum, Norway.
Key points
From the paper's conclusion- Although the cuffless device shows promise in successfully tracking nocturnal BP, full validation is required before clinical use.
Abstract
BACKGROUND: Cuffless blood pressure (BP) devices have demonstrated potential in estimating daytime BP but have not been able to predict nocturnal BP precisely, creating difficulties when the intended use is ambulatory BP monitoring (ABPM). We evaluated the ability of a photoplethysmography-based cuffless BP device, with the intended use 24-hour ABPM, to estimate nocturnal BP compared to a cuff-based oscillometric device. METHODS: Participants underwent 24-hour ABPM with the cuffless device placed on the non-dominant upper arm and a validated cuff-based oscillometric device on the dominant arm. The difference between mean awake and asleep BP (awake-asleep BP change) was calculated for both devices and used in the analyses. The statistical analyses were guided by the Awake/Asleep test (European Society of Hypertension 2023 Recommendations), however, there were methodological deviances from these recommendations. Most importantly, we used only cuffless BP values that were obtained simultaneously with the reference device. RESULTS: Data from 50 participants, aged 18-80 years, with mean (standard deviation; SD) inclusion office systolic/diastolic BP (SBP/DBP) 120.1 (11.5)/75.3 (7.9), were analyzed. The cuffless device achieved a mean difference [95% limits of agreement] of 1.6 [-8.0, 11.2] mmHg for SBP and 1.2 [-6.7, 9.1] mmHg for DBP in awake-asleep BP changes compared to the reference. Pearson correlation coefficients (95% confidence interval) for SBP/DBP were 0.75 (0.59, 0.85)/0.74 (0.58, 0.84). Our population had an overrepresentation of normal and low BP. CONCLUSION: Although the cuffless device shows promise in successfully tracking nocturnal BP, full validation is required before clinical use.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Continual cuffless blood pressure monitoring in children: agreement and feasibility compared with ambulatory blood pressure monitoring.Journal of hypertension · 2026
- Real-world performance of a ring-type cuffless blood pressure device compared with ambulatory blood pressure monitoring.Clinical hypertension · 2026
- Validation of ring-type cuffless blood pressure (BP) monitoring device for detecting awake-asleep BP variations compared to ambulatory BP monitoring device.Hypertension research : official journal of the Japanese Society of Hypertension · 2026
- Long-Term Stability and Blood Pressure Change Tracking with a Cuffless Wearable Device in Hypertensive Patients.Blood pressure · 2026
- Evaluation of Alysis-001 cuffless blood pressure estimation algorithm against the European Society of Hypertension awake/asleep test criteria in hypertensive patients.Hypertension research : official journal of the Japanese Society of Hypertension · 2026
- Validation study on cuffless ambulatory blood pressure monitoring method using bio-impedance, impedance cardiography and electrocardiography.European journal of applied physiology · 2026