Optimising electron beam 6 MeV radiotherapy: field-size-dependent dosimetric impact of perforated thermoplastic immobilisation masks.
Mohammed Rezzoug, Yassine Oulhouq, Omar Hamzaoui, Mustapha Zerfaoui, Abdeslem Rrhioua, Dikra Bakari
LPPMOIA, Faculty of Sciences, Mohammed First University, Oujda, Morocco. · LaRSA, National School of Applied Sciences, Mohammed First University, Oujda, Morocco.
Key points
From the paper's conclusion- Masks with 4 mm perforations consistently outperformed 2 mm variants in dosimetric accuracy while maintaining effective immobilisation. Tailored mask selection based on field size improves clinical outcomes in electron beam radiotherapy.
Abstract
BACKGROUND: Thermoplastic immobilisation masks are widely used in head and neck radiotherapy to ensure patient stability and reproducible positioning during treatment. Because these masks are placed directly in the beam path and consist of perforated polymer materials, they may influence electron dose distributions through additional attenuation and scattering effects, particularly for low-energy electron beams. This study quantifies the dosimetric effects of perforated thermoplastic masks, considering different hole diameters and electron field sizes. MATERIALS AND METHODS: Monte Carlo simulations using GATE modelled 6 MeV electron beams passing through thermoplastic masks with 2 mm and 4 mm hole diameters across 3 × 3 cm2, 6 × 6 cm2, and 10 × 10 cm2 fields. Surface dose (D0), depth of maximum dose (R100), isodose depths (R90, R50), penumbra, flatness and gamma analysis (2%/2 mm and 3%/3 mm) were evaluated. RESULTS: Masks increased D0 up to +5.9% for 2 mm holes and +4.18% for 4 mm holes. R100 showed up to -21.43% reduction for 2 mm masks, with lesser shifts using 4 mm holes. Penumbra widening was also detected, with a maximum relative increase of +4.40% at Zmax for the 6 × 6 cm2 field using the 2 mm mask. Gamma pass rates were better for 4 mm masks, especially at shallow depths. CONCLUSIONS: Masks with 4 mm perforations consistently outperformed 2 mm variants in dosimetric accuracy while maintaining effective immobilisation. Tailored mask selection based on field size improves clinical outcomes in electron beam radiotherapy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Impact of magnetic field orientation and strength on dose distributions in 6 MV photon beams: A Monte Carlo study.Radiation and environmental biophysics · 2026
- Evaluation of the environ SRS immobilization mask system for HyperArc: Dosimetric attenuation, dose calculation accuracy, and clinical setup reproducibility.Journal of applied clinical medical physics · 2026
- Maskless head and neck radiotherapy using surface-guided radiation therapy: initial clinical feasibility of real-time setup, intrafraction monitoring, and automatic beam-hold.Journal of applied clinical medical physics · 2026
- A Compact Eight-Parameter Analytical Model of Photon-Beam Lateral Dose Profiles: Plateau, Penumbra and Out-of-Field Tail Decomposition for Elekta Linear Accelerators.Biomedical physics & engineering express · 2026
- Dosimetric assessment of radiation fields up to 5 × 5 cm 2 and architectural optimization of a diode-based 3D Detector for radiotherapyJournal of Physics: Conference Series · 2026
- Characterization and Mitigation of Boundary Sampling Artifacts in Electron Phase Space Generation for Clinical FLASH-RT Deployment.ArXiv · 2026