A Compact Eight-Parameter Analytical Model of Photon-Beam Lateral Dose Profiles: Plateau, Penumbra and Out-of-Field Tail Decomposition for Elekta Linear Accelerators.
Mustapha Zerfaoui, Ibrahim Ait Ouaggou, Mohammed Rezzoug, Omar Hamzaoui, Yassine Oulhouq, Samir Didi, Abdeslem Rrhioua, Dikra Bakari,
physique, Laboratory of Physics of Matter and Radiation Faculty of Sciences, Mohammed first University Oujda, Morocco, UMP, Oujda, 60000, Morocco. · Mohammed First University Oujda, ibrahim.aitouaggou@ump.ac.ma, Oujda, 60000, Morocco.
Key points
From the paper's conclusion- Each parameter retains a direct physical meaning, and the parameters recovered simulated jaw-calibration and steering errors set at or below the AAPM TG-142 tolerances, supporting their use in beam-data QA and commissioning; independent dose calculation will require a continuous parameterization in depth and field size.
Abstract
To develop and validate a compact analytical model of megavoltage photon-beam lateral dose profiles, with eight physically interpretable parameters per profile, for beam-data quality assurance (QA) and treatment planning system (TPS) commissioning support. The relative dose at off-axis position x, for a field of half-width R, is expressed as D(x,R) = A·F(x,R)·P(x,R) + S(x,R), where F (a fourth-order polynomial in x/R) describes the plateau, P (an error-function pair of width σ) the penumbra, and S (an exponential) the out-of-field tail. The parameters were fitted to 58 in-plane profiles measured on an Elekta linac at 6, 10 and 15 MV (1×1 to 40×40 cm², depths 1.5-20 cm), and the model was tested on 14 independent profiles (second linac, Monte Carlo). Agreement was assessed with the root-mean-square error (RMSE), R² and the global gamma index at 2%/2 mm and 1%/1 mm. All profiles passed 2%/2 mm at 100%, with a mean RMSE of 0.551% (range 0.24-1.19%) and R² ≥ 0.9978; the independent profiles gave a mean RMSE of 0.541%. At 1%/1 mm, 99.2% of points passed, and all 6 MV profiles and all fields ≥ 4×4 cm² passed at ≥ 95%; the remaining deviations were confined to small 10-15 MV fields and low-gradient regions. The parameters followed physically consistent trends: σ increased with depth and, for fields ≥ 5×5 cm², with energy, B increased with depth and with field size up to 30×30 cm², α decreased with depth, and H and H₄ tracked the loss of horns with depth. Each parameter retains a direct physical meaning, and the parameters recovered simulated jaw-calibration and steering errors set at or below the AAPM TG-142 tolerances, supporting their use in beam-data QA and commissioning; independent dose calculation will require a continuous parameterization in depth and field size.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Beam data and model validation for Elekta SRS cone collimator in 6MV and 6MVFFF beams.Medical dosimetry : official journal of the American Association of Medical Dosimetrists · 2026
- Pulse-width-specific phase space modeling and reference beam characterization for UHDR electron LINAC in FLASH-RTPhysics in Medicine and Biology · 2026
- Impact of magnetic field orientation and strength on dose distributions in 6 MV photon beams: A Monte Carlo study.Radiation and environmental biophysics · 2026
- Toward dynamic Monte Carlo source modeling: measurement variability and primary electron source-width sensitivity in very small photon fields.Frontiers in oncology · 2026
- VHEE beam parameter extraction from depth-dose and lateral profile measurements: a Monte Carlo parametric study.Radiological physics and technology · 2026
- Evaluation of GPU-accelerated Monte Carlo beam and agility MLC modeling in the Monaco treatment planning system.Journal of applied clinical medical physics · 2026