Impact of magnetic field orientation and strength on dose distributions in 6 MV photon beams: A Monte Carlo study.
Mohammed Rezzoug, Yassine Oulhouq, Omar Hamzaoui, Mustapha Zerfaoui, Abdeslem Rrhioua, Dikra Bakari
Laboratory of Theoretical Physics, Particles, Modeling and Energies, Mohammed First University, Faculty of Science, Boulevard Mohamed VI, BP 717, Oujda-Angad, 60000, Morocco. med.rezzoug@ump.ac.ma. · Laboratory of Theoretical Physics, Particles, Modeling and Energies, Mohammed First University, Faculty of Science, Boulevard Mohamed VI, BP 717, Oujda-Angad, 60000, Morocco.
Key points
From the paper's conclusion- This study investigates the influence of magnetic field orientation (Ox, Oy, Oz) and strength (0-3 T) on percentage depth dose (PDD) and dose profile parameters for 6 MV clinical photon beams using Monte Carlo simulations.
Abstract
This study investigates the influence of magnetic field orientation (Ox, Oy, Oz) and strength (0-3 T) on percentage depth dose (PDD) and dose profile parameters for 6 MV clinical photon beams using Monte Carlo simulations. Monte Carlo simulations were performed using GATE (version 9.2) with a validated 6 MV phase space source. Three field sizes (10 × 10, 15 × 15, and 20 × 20 cm²) were evaluated under magnetic fields ranging from 0 to 3 T in increments of 0.5 along three orthogonal directions (Ox, Oy, Oz). PDD parameters including R100, D100, D200, surface dose (Ds), and quality index (Qi) were analyzed. Dose profile characteristics including field size and penumbra width were evaluated in both X and Y directions. Magnetic field orientation and strength significantly affected dose distributions. For Ox and Oy orientations, R100 decreased progressively with increasing field strength: from 16 to 17 mm at 0 T to 7-9 mm at 3 T across all field sizes. Surface dose increased substantially from 40 to 45% at 0 T to 58-65% at 3 T. For Oz orientation, R100 showed minimal variation (10-13 mm across all field strengths), while surface dose increased more substantially to 56-66%. Penumbra width and field size demonstrated orientation-dependent variations, with minimal changes for Ox and Oy, but systematic reductions for Oz orientation. It is concluded that magnetic field presence substantially alters dose deposition patterns in an orientation-dependent manner. These findings are critical for MRI-guided radiotherapy systems and require specific dosimetric considerations for treatment planning and quality assurance protocols.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Compact Eight-Parameter Analytical Model of Photon-Beam Lateral Dose Profiles: Plateau, Penumbra and Out-of-Field Tail Decomposition for Elekta Linear Accelerators.Biomedical physics & engineering express · 2026
- Magnetic-Field-Induced Secondary-Electron Effects in MR-Guided Radiotherapy: From Dosimetric Mechanisms to Workflow-Based Clinical Risk Management.Technology in cancer research & treatment · 2026
- Magnetically enhanced peak‐to‐valley dose ratio for megavoltage photon minibeam radiotherapyMedical Physics · 2026
- Pulse-width-specific phase space modeling and reference beam characterization for UHDR electron LINAC in FLASH-RTPhysics in Medicine and Biology · 2026
- Dosimetric comparison of 6-MV flattened and flattening filter-free photon beams using Monte Carlo simulationNuclear Engineering and Technology · 2026
- Beam data and model validation for Elekta SRS cone collimator in 6MV and 6MVFFF beams.Medical dosimetry : official journal of the American Association of Medical Dosimetrists · 2026