Dosimetric validation of biaxially rotational dynamic-radiation therapy using an O-ring linear accelerator.
Takahiro Aoyama, Tomoki Kitagawa, Yutaro Koide, Yasunori Ishiguro, Ryoma Tomoda, Shingo Hashimoto, Hiroyuki Tachibana, Takeshi Kodaira
Department of Radiation Oncology, Aichi Cancer Center Hospital, Nagoya, Japan.
Key points
From the paper's conclusion- This study demonstrated dosimetric validation of a biaxially rotating O-ring linear accelerator using TG-119 and TG-244 test cases. The system achieved high accuracy in both cVMAT and DSA deliveries, supporting the dosimetric readiness for clinical implementation.
Abstract
BACKGROUND: Biaxially rotating O-ring linear accelerators equipped with biaxially rotating gantry-ring mechanics enable non-coplanar arc delivery without the need for couch rotation. This architecture is intended to achieve conformal dose distributions and enable efficient non-coplanar delivery compared with couch-rotated techniques. However, clinical implementation requires dosimetric verification to confirm delivery accuracy. This study aimed to validate the dosimetric accuracy and treatment efficiency of a biaxially rotating O-ring linac using AAPM Task Group (TG)-119 and TG-244 test cases, with reference to TG-218 acceptance criteria. MATERIALS AND METHODS: In this retrospective phantom-based dosimetric validation study, 26 treatment plans were generated in RayStation based on TG-119 and TG-244 and delivered as coplanar VMAT (cVMAT) or non-coplanar VMAT [dynamic swing arc (DSA)]. Point dose measurements were performed with an A1SL ionisation chamber, and dose distribution verification was conducted using the Delta4 Phantom+. Monitor units (MU) and beam-on time (BOT) were compared between cVMAT and DSA. RESULTS: All plans met the ±3% tolerance for absolute point-dose error (measured vs. calculated), with a mean error of -0.21% ± 1.09%. Gamma analysis (3%/2 mm, 10% threshold) showed a mean pass rate of 99.3%, exceeding the 95% threshold. Significant differences were observed in MU and BOT between cVMAT (median 862.4 MU, 104.0 s) and DSA (median 740.5 MU, 136.5 s). CONCLUSIONS: This study demonstrated dosimetric validation of a biaxially rotating O-ring linear accelerator using TG-119 and TG-244 test cases. The system achieved high accuracy in both cVMAT and DSA deliveries, supporting the dosimetric readiness for clinical implementation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Off-axis Winston-Lutz quality assurance for static and rotational delivery on a gimbal-mounted O-ring linear accelerator.Biomedical physics & engineering express · 2026
- Comparative Analysis of Arc Geometrical Configurations and Jaw Tracking for Peripheral Low-Dose Bath in Whole Breast VMAT.Technology in cancer research & treatment · 2026
- Commissioning and clinical positioning assessment of a novel surface-guided radiation therapy (SGRT) system at a C-Arm linear accelerator.Journal of applied clinical medical physics · 2026
- RapidArc dynamic versus IMRT and VMAT for endometrial cancer SIB radiotherapy.Radiation oncology (London, England) · 2026
- Clinical implementation and performance evaluation of a motorized MR-compatible OCTAVIUS 4D system for SBRT patient-specific QA on a 1.5 T MR-LINAC.Journal of applied clinical medical physics · 2026
- ExacTrac Dynamic Surface-Guided Setup Versus Tattoo-Based Positioning in Breast Cancer Radiotherapy: A Retrospective Comparison of Setup Accuracy and Preparation Time.Cancers · 2026