Off-axis Winston-Lutz quality assurance for static and rotational delivery on a gimbal-mounted O-ring linear accelerator.
Takahiro Aoyama, Maho Morishita, Tomoki Kitagawa, Ryoma Tomoda, Yasunori Ishiguro, Yutaro Koide, Takeshi Kodaira
Department of Radiation Oncology, Aichi Cancer Center, 1-1, Kanokoden, Chikusa-ku, Nagoya, Aichi, 464-8681, Japan. · Aichi Cancer Center, 1-1, Kanokoden, Chikusa-ku, Naogoya, Aichi, 464-8681, Japan.
Key points
From the paper's conclusion- We assessed a commercial Winston-Lutz (WL) and off-axis Winston-Lutz (OAWL) procedure for field-to-target coincidence at the machine isocenter and at ±30 mm along the superior-inferior axis.
Approach: A MultiMet-WL Cube phantom and DoseLab Pro were used to analyze cine electronic portal imaging device images acquired during static four-port irradiation, 360° gantry rotations, and a BROAD-RT tr
Abstract
OXRAY is a gimbal-mounted O-ring linear accelerator that can shift the radiation isocenter away from the machine isocenter and deliver biaxial rotational dynamic radiation therapy (BROAD-RT). Published commissioning studies have characterized dynamic beam positioning with in-house analysis, but static gimbal-shifted positions have not been evaluated with a commercially available quality assurance system. We assessed a commercial Winston-Lutz (WL) and off-axis Winston-Lutz (OAWL) procedure for field-to-target coincidence at the machine isocenter and at ±30 mm along the superior-inferior axis. Approach: A MultiMet-WL Cube phantom and DoseLab Pro were used to analyze cine electronic portal imaging device images acquired during static four-port irradiation, 360° gantry rotations, and a BROAD-RT trajectory. Main results: Static 2D displacement was 0.38 ± 0.20 mm (maximum 0.73 mm) at the machine isocenter, 0.45 ± 0.15 mm (0.70 mm) at +30 mm, and 0.45 ± 0.20 mm (0.79 mm) at -30 mm. During gantry rotation, mean displacements were 0.25-0.34 mm and the maximum was 0.68 mm. BROAD-RT yielded 0.32 ± 0.16 mm with a maximum of 0.65 mm. Clockwise and counterclockwise rotations showed similar angle-dependent profiles within the measurement session. The maximum recorded-minus-commanded gimbal tilt residual was 0.004°, showing high agreement within the control system. All measured values were below 1 mm. Significance: This commercial procedure directly tests static gimbal-shifted positions that are not covered by center-only WL testing and provides the same angle-resolved endpoint for rotational and BROAD-RT delivery without custom phantom hardware or analysis software.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Dosimetric validation of biaxially rotational dynamic-radiation therapy using an O-ring linear accelerator.Reports of practical oncology and radiotherapy : journal of Greatpoland Cancer Center in Poznan and Polish Society of Radiation Oncology · 2026
- Commissioning and clinical positioning assessment of a novel surface-guided radiation therapy (SGRT) system at a C-Arm linear accelerator.Journal of applied clinical medical physics · 2026
- A patient-specific Winston-Lutz system for multi-target, single isocentre radiotherapy.Biomedical physics & engineering express · 2026
- Optimizing TrueBeam jaw calibrations to reduce positional errors at customizable gantry and collimator anglesJournal of Applied Clinical Medical Physics · 2026
- Clinical implementation and performance evaluation of a motorized MR-compatible OCTAVIUS 4D system for SBRT patient-specific QA on a 1.5 T MR-LINAC.Journal of applied clinical medical physics · 2026
- ExacTrac Dynamic Surface-Guided Setup Versus Tattoo-Based Positioning in Breast Cancer Radiotherapy: A Retrospective Comparison of Setup Accuracy and Preparation Time.Cancers · 2026