Sandwich liner fracture after 21 years from revision total hip arthroplasty: a case report.
Yerl Bo Sung, Hyungtae Kim, Juyoung Jeong
Department of Orthopedic surgery Namyangju Hyundai Hospital. · Department of Orthopedic Surgery Inje University Sanggye Paik Hospital.
Key points
From the paper's conclusion- Long-term surveillance and prompt evaluation of mechanical symptoms are advised for older-generation sandwich liners.
Abstract
INTRODUCTION: Ceramic-polyethylene sandwich liners were introduced to reduce fracture, but high failure rates were reported. We report a fracture clinically recognized 21 years after revision total hip arthroplasty (THA). CASE PRESENTATION: A 64-year-old man presented with two weeks of hip pain, clicking, and gait disturbance without trauma. He underwent primary THA for secondary coxarthrosis after slipped capital femoral epiphysis and revision THA for aseptic loosening three years later. Radiographs demonstrated eccentric femoral-head migration and radiopaque fragments. Computed tomography confirmed ceramic layer fracture and migration into the obturator internus muscle. MANAGEMENT AND OUTCOMES: Re-revision of the acetabular cup with a fourth-generation ceramic articulation was performed, with the well-fixed femoral stem retained. At 43 months, he returned to recreational hiking; the modified Harris Hip Score improved from 42.9 to 87.5. Residual fragments remained stable. CONCLUSION: Long-term surveillance and prompt evaluation of mechanical symptoms are advised for older-generation sandwich liners.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Catastrophic acetabular wear following polyethylene liner dissociation with locking-tab damage after total hip arthroplasty: a case report.Frontiers in surgery · 2026
- Ceramic component fracture following trauma in a bilateral total hip arthroplasty.Orthopedic reviews · 2026
- Early Catastrophic Failure of Highly-Cross-Linked Polyethylene Liner Due to Acetabular Shell Screw Placement: A Case Report.Cureus · 2026
- Patterns of wear, fracture, and re-revision in ceramic-on-ceramic total hip arthroplasty: a 31-year retrieval study.Hip international : the journal of clinical and experimental research on hip pathology and therapy · 2026
- Revision total hip arthroplasty for osteolysis and metallosis 17 years post-THA: a case report and literature review.International journal of surgery case reports · 2026