Catastrophic Failure of a Prosthetic Femoral Stem due to Component Malposition: A Case Report.
Adam C Pasquinelly, Brandon Sharkey, Alicia M Nahhas, Isaac A Arefi, David Yatsonsky, Connor D Burke, Gregory M Georgiadis
University of Toledo College of Medicine and Life Sciences, Toledo, Ohio, USA, utoledo.edu. · Department of Orthopaedic Surgery, University of Toledo Medical Center, Toledo, Ohio, USA, utoledo.edu.
Key points
From the paper's conclusion- We present the case of a patient with a history of prior acetabular fracture who underwent conversion to a dual mobility (DM) THA and subsequently fractured at the neck of the femoral component 7 years after surgery.
Abstract
Implant fracture is a rare but devastating complication of total hip arthroplasty (THA). We present the case of a patient with a history of prior acetabular fracture who underwent conversion to a dual mobility (DM) THA and subsequently fractured at the neck of the femoral component 7 years after surgery. At revision, both the femoral and acetabular components were noted to be clinically retroverted, suggesting chronic anterior impingement as the cause of the catastrophic failure.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Catastrophic Acetabular Osteolysis and Component Migration Secondary to Surgical Cup Malpositioning: A Case Report.Cureus · 2026
- Case Report: One-stage comprehensive reconstruction for intrapelvic cup migration with Paprosky IIIB acetabular bone loss and Vancouver B3 periprosthetic femoral fracture after total hip arthroplastyFrontiers in Surgery · 2026
- Distal femur replacement after catastrophic failure of a total knee arthroplasty.BMJ case reports · 2026
- Transcutaneous extrusion of a migrated bipolar hip hemiarthroplasty 9 years after implantation: a case report.Journal of surgical case reports · 2026
- Ceramic component fracture following trauma in a bilateral total hip arthroplasty.Orthopedic reviews · 2026
- Early Catastrophic Failure of Highly-Cross-Linked Polyethylene Liner Due to Acetabular Shell Screw Placement: A Case Report.Cureus · 2026