Catastrophic acetabular wear following polyethylene liner dissociation with locking-tab damage after total hip arthroplasty: a case report.

Weida Li, Zhixing Liang, Zhuangzhuang Dong, Qian Sun, Yu Hou

Department of Joint Surgery, Affiliated Hospital of Hebei University, Baoding, China.

Case report

Key points

From the paper's conclusion
  • Conclusion This case illustrates the potential for clinically occult progression from liner dissociation to catastrophic head-on-shell wear.

Abstract

BACKGROUND: Polyethylene liner dissociation is an uncommon but potentially severe complication of total hip arthroplasty (THA). Loss of the intended bearing interface can permit direct femoral-head-acetabular-shell articulation, with progressive shell wear, metallic debris deposition, and acetabular bone loss. CASE PRESENTATION: A 73-year-old woman developed left hip pain after an accidental sprain in February 2025, having reported no hip-related symptoms after primary THA in 2020. The index Mindray implants included a 48-mm acetabular shell, a highly cross-linked polyethylene liner, a 28-mm femoral head, and a size-10 stem. Radiographs and CT with three-dimensional volume-rendered reconstruction were used to assess acetabular bone loss, classified as Paprosky type IIA. Revision revealed polyethylene liner dissociation, catastrophic shell wear-through, extensive black metallic debris deposition, and a well-fixed femoral stem. Partial locking-tab damage was identified on gross inspection, but formal retrieval analysis was not performed. Acetabular revision with iliac-crest autografting was performed while retaining the femoral stem. At 6 months, the Harris Hip Score improved from 46 to 82, and radiographs showed a stable acetabular component with evidence of graft incorporation. CONCLUSION: This case illustrates the potential for clinically occult progression from liner dissociation to catastrophic head-on-shell wear. Gross locking-tab damage did not establish the initiating failure mechanism. Component-preserving acetabular revision was associated with a favorable short-term clinical and radiographic outcome in this patient.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Catastrophic acetabular wear following polyethylene liner dissociation with locking-tab damage after total hip arthroplasty: a case report — Science Skim