Catastrophic acetabular wear following polyethylene liner dissociation with locking-tab damage after total hip arthroplasty: a case report.
Weida Li, Zhixing Liang, Zhuangzhuang Dong, Qian Sun, Yu Hou
Department of Joint Surgery, Affiliated Hospital of Hebei University, Baoding, China.
Key points
From the paper's conclusion- Conclusion This case illustrates the potential for clinically occult progression from liner dissociation to catastrophic head-on-shell wear.
Abstract
BACKGROUND: Polyethylene liner dissociation is an uncommon but potentially severe complication of total hip arthroplasty (THA). Loss of the intended bearing interface can permit direct femoral-head-acetabular-shell articulation, with progressive shell wear, metallic debris deposition, and acetabular bone loss. CASE PRESENTATION: A 73-year-old woman developed left hip pain after an accidental sprain in February 2025, having reported no hip-related symptoms after primary THA in 2020. The index Mindray implants included a 48-mm acetabular shell, a highly cross-linked polyethylene liner, a 28-mm femoral head, and a size-10 stem. Radiographs and CT with three-dimensional volume-rendered reconstruction were used to assess acetabular bone loss, classified as Paprosky type IIA. Revision revealed polyethylene liner dissociation, catastrophic shell wear-through, extensive black metallic debris deposition, and a well-fixed femoral stem. Partial locking-tab damage was identified on gross inspection, but formal retrieval analysis was not performed. Acetabular revision with iliac-crest autografting was performed while retaining the femoral stem. At 6 months, the Harris Hip Score improved from 46 to 82, and radiographs showed a stable acetabular component with evidence of graft incorporation. CONCLUSION: This case illustrates the potential for clinically occult progression from liner dissociation to catastrophic head-on-shell wear. Gross locking-tab damage did not establish the initiating failure mechanism. Component-preserving acetabular revision was associated with a favorable short-term clinical and radiographic outcome in this patient.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Early Catastrophic Failure of Highly-Cross-Linked Polyethylene Liner Due to Acetabular Shell Screw Placement: A Case Report.Cureus · 2026
- Sandwich liner fracture after 21 years from revision total hip arthroplasty: a case report.Orthopedic reviews · 2026
- Catastrophic Acetabular Osteolysis and Component Migration Secondary to Surgical Cup Malpositioning: A Case Report.Cureus · 2026
- Catastrophic Failure of a Prosthetic Femoral Stem due to Component Malposition: A Case Report.Case reports in orthopedics · 2026
- Computational methodology for the evaluation of tangential stresses and torsional moments in heterogeneous total hip arthroplasty assembliesJournal of Physics: Conference Series · 2026
- Revision total hip arthroplasty for osteolysis and metallosis 17 years post-THA: a case report and literature review.International journal of surgery case reports · 2026