Surface alterations of the MicroMega One RECI reciprocating instrument under two kinematics: A scanning electron microscopy evaluation.

Kevin Denis Lavado-Cuicapusa, Maria Mihaela Iuga, Carlos Mendiola-Aquino, Antonio Ernesto Denegri-Hacking, Carlos Garcia-Puente, Alejandro Jaime, Mario Casaretto-Gamonal

Department of Stomatological Clinic, Cayetano Heredia Peruvian University, Lima, Perú. · Faculty of Health Sciences, Private University of Tacna, Tacna, Perú.

Key points

From the paper's conclusion
  • Within the limitations of this ex vivo paired pre-post study, repeated use increased surface alterations in One RECI instruments. ARI showed a tendency toward fewer defects than NOR, although the between-kinematic comparison should be interpreted cautiously because of the exploratory sample size. These findings support limiting instrument reuse and suggest that adaptive reciprocation may reduce cumulative surface damage.

Abstract

BACKGROUND: This ex vivo study compared surface alterations of MicroMega One RECI 25/.06 instruments after use under two reciprocating kinematics: Adaptive Intelligent Reciprocation (ARI) and the manufacturer's standard reciprocating motion (NOR). MATERIALS AND METHODS: Twenty new One RECI instruments were first examined by scanning electron microscopy (SEM) to establish their factory-new baseline condition (FAB). The same instruments were then randomly allocated to ARI (n=10) or NOR (n=10). Each file was used to prepare three mesial canals of extracted mandibular molars with moderate curvature and was subsequently re-examined by SEM. Surface defects were scored by instrument segment and classified according to predefined operational criteria as dull edges, deformed flutes, microcracks, craters, or tip deformation. Data were analyzed using descriptive statistics, non-parametric tests, and Poisson regression modeling. RESULTS: Dull edges and deformed flutes were the most frequently observed defects. Factory-new instruments showed minimal baseline alterations, whereas used instruments exhibited significantly higher defect counts. In the unadjusted pairwise comparison, ARI and NOR did not differ significantly. However, Poisson regression suggested a higher expected number of defects in the NOR group compared with ARI. CONCLUSIONS: Within the limitations of this ex vivo paired pre-post study, repeated use increased surface alterations in One RECI instruments. ARI showed a tendency toward fewer defects than NOR, although the between-kinematic comparison should be interpreted cautiously because of the exploratory sample size. These findings support limiting instrument reuse and suggest that adaptive reciprocation may reduce cumulative surface damage.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers