Clinical Surface Alteration of Reciprocating Nickel-Titanium Instruments Following Root Canal Preparation and Sterilization: A Three-Dimensional FE-SEM-Based Analysis.
Nguyen-Huu-Phuoc Huynh, Hong-Hai Le, Nguyen-Minh-Hieu Tran, Hoang-Vinh Le, Huynh-Anh Bui, Thi-Thuy-Trang Huynh, Hoang-Lan-Anh Le, Phuong-Doan Phan,
Faculty of Dentistry, University of Medicine and Pharmacy at Ho Chi Minh City, Ho Chi Minh City, Vietnam, hcmut.edu.vn. · Faculty of Odonto-Stomatology, University of Health Sciences, Vietnam National University, Ho Chi Minh City, Vietnam, vnu.edu.vn.
Key points
From the paper's conclusion- Routine clinical use and sterilization can induce detectable surface modifications in reciprocating NiTi instruments. Three-dimensional FE-SEM analysis provides a clinically relevant approach for assessing functional surface evolution under actual treatment conditions.
Abstract
BACKGROUND: The study aims to evaluate the surface alteration of the reciprocating nickel-titanium (NiTi) instrument after clinical instrumentation. METHODS: This clinical investigation evaluated surface changes of reciprocating NiTi instruments after routine endodontic use and sterilization. Eleven One RECI instruments, each used to prepare a single molar canal in one patient (11 patients, 11 canals, and 11 instruments), were evaluated. Maxillary and mandibular molars requiring primary treatment were instrumented using One RECI files under standardized clinical protocols. Following use-and-steam autoclave sterilization, instruments were analyzed using field-emission scanning electron microscopy (FE-SEM). Paired stereoscopic images obtained at standardized locations were reconstructed into three-dimensional surface models using Mountains software (Digital Surf, Besancon, France). Areal surface roughness parameters, including Sa, Sq, and Sz, were calculated according to ISO 25178. RESULTS: Clinical use resulted in increased Sa and Sq values at the cutting edge of the instruments, indicating measurable surface alteration after routine treatment and sterilization. These findings were consistent with trends previously reported under preclinical conditions. CONCLUSION: Routine clinical use and sterilization can induce detectable surface modifications in reciprocating NiTi instruments. Three-dimensional FE-SEM analysis provides a clinically relevant approach for assessing functional surface evolution under actual treatment conditions.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Effects of Different Irrigation Protocols on the Surface Morphology and Elemental Composition of Super One File Nickel-Titanium Instruments.Bioengineering (Basel, Switzerland) · 2026
- Surface alterations of the MicroMega One RECI reciprocating instrument under two kinematics: A scanning electron microscopy evaluation.Journal of clinical and experimental dentistry · 2026
- Effect of clinical use and autoclave sterilization on the dynamic cyclic fatigue resistance of ProTaper Ultimate instruments.Odontology · 2026
- The impact of different autoclaving cycles on the properties of rotary endodontic files: an in vitro study.BMC oral health · 2026
- Effect of Sodium Hypochlorite Exposure on Cyclic Fatigue Resistance of Two Heat-Treated Nickel-Titanium Rotary Instruments: An In Vitro Study.Cureus · 2026
- Effect of Intermittent Cleaning on the Longevity of Nickel– Titanium Rotary Files During Root Canal Instrumentation – An In-Vitro StudyInternational Journal of Drug Delivery Technology · 2026