Development of a low-cost 3D-printed phantom for end-to-end testing of model-based dose calculation algorithms (MBDCAs) in high-dose-rate (HDR) brachytherapy.
Sean Grace, Lin Wang, Yijian Cao, Lyu Huang
Department of Physics and Astronomy, Hofstra University, Hempstead, New York, USA. · Northwell, New Hyde Park, New York, USA.
Key points
From the paper's conclusion- This low-cost, modular, 3D-printed phantom provides a practical platform for E2E validation of MBDCAs under controlled heterogeneous conditions. Its customizable design may complement existing commissioning methods, particularly in resource-limited settings.
Abstract
BACKGROUND: Model-Based Dose Calculation Algorithms (MBDCAs), such as AcurosBV, have improved dosimetric accuracy in high-dose-rate (HDR) brachytherapy by accounting for medium heterogeneities and variations in scatter conditions. However, their safe and effective clinical implementation requires thorough commissioning, of which end-to-end (E2E) validation using physical phantoms is an important component. Commercial anthropomorphic phantoms can be costly and may offer limited flexibility for customized testing configurations. PURPOSE: This study aims to design, fabricate, and evaluate a novel low-cost, modular, 3D-printed phantom for E2E validation of MBDCAs in HDR brachytherapy. The phantom was developed to provide a practical, measurement-based tool for commissioning and quality assurance under controlled heterogeneous conditions, particularly in resource-limited settings. METHODS: A modular phantom incorporating heterogeneity plugs (moderate-density, air cavity, and calcium carbonate) was 3D printed using Polylactic acid (PLA) on a Bambu X1 Carbon printer. EBT4 films were scanned on an Epson 10000XL flatbed scanner, calibrated and used for planar dose measurements. The phantom was scanned with a Siemens CT simulator. Treatment plans were generated using BrachyVision in Eclipse v16.1, with dose distributions calculated using both TG-43 and AcurosBV algorithms. The plan was delivered with a GammaMed Plus iX afterloader. Film measurements were compared against calculated dose maps using gamma analysis and dose profile comparisons. RESULTS: The 3D-printed phantom demonstrated high dimensional fidelity (± 0.15 mm) and allowed for reproducible catheter placement and film positioning. AcurosBV calculations showed close agreement with film measurements, with gamma passing rates of 95% at 2%/2 mm criterion and 94.2% at 3%/1 mm criterion. Mean dose differences between AcurosBV and film were 1.7% ± 1.9% for the vertical profile and -0.5% ± 1.7% for the horizontal profile. Under the controlled heterogeneous configuration of the 3D-printed phantom, TG-43 predicted higher doses than film by 3.6% ± 1.2% and 8.2% ± 1.5% for the vertical and horizontal profiles, respectively. These differences reflect the combined influence of the non-water-equivalent phantom body and the incorporated heterogeneous inserts. CONCLUSIONS: This low-cost, modular, 3D-printed phantom provides a practical platform for E2E validation of MBDCAs under controlled heterogeneous conditions. Its customizable design may complement existing commissioning methods, particularly in resource-limited settings.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Clinical test cases for commissioning, QA, and benchmarking of model‐based dose calculation algorithms in¹ ⁹² Ir HDR gynecologic tandem and ring brachytherapyMedical Physics · 2026
- Preliminary feasibility results for implementing a phantom model for external auditing of HDR brachytherapy with an Ir-192 HDR source: an IAEA collaborative project.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Verifying calculated dose to the breast in stereotactic radiotherapy: A practical phantom-based method.Journal of applied clinical medical physics · 2026
- Development and validation of a 3D-printed dosimetry phantom for paediatric radiologyPhysical and Engineering Sciences in Medicine · 2026
- Investigation of TG-43 dose calculation algorithm for volumetric HDR Brachytherapy using Monte Carlo simulation in TOPAS.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- A novel 3D-printed applicator for image-guided interstitial brachytherapy in locally advanced cervical cancer: superior dosimetry, efficacy, and safetyRadiation Oncology · 2026