Key points
From the paper's conclusion- The 4R-TT/CBT configuration provided superior biomechanical performance in lumbar PSO, enhancing construct stiffness, and substantially reducing stresses on rods and screws compared to the 2R-TT and 4R-TT configurations. However, this was a FE study with inherent simplifications. Therefore, the findings should be further confirmed by cadaveric or clinical studies in the future.
Abstract
INTRODUCTION: Although multi-rods constructs (Multi-RCs) have been demonstrated to reduce the risk of rod fracture (RF) following pedicle subtraction osteotomy (PSO), they offer limited improvement in terms of screw-related complications. The double-trajectory technique (combining traditional trajectory (TT) and cortical bone trajectory (CBT) screws) has been demonstrated to enhance screw fixation. Consequently, the employment of the double-trajectory technique in conjunction with Multi-RCs has the capacity to augment the fixation strength of both rods and screws. Thus, the study aimed to evaluate the biomechanical performance of a novel configuration combining double-trajectory screws with multi-rod in lumbar PSO. METHODS: A validated L1-5 nonlinear FE model was modified to simulate a 25° PSO at L3. Three configurations were analyzed under both normal and osteoporotic conditions: (a) 2R-TT (one rod with TT screws on each side); (b) 4R-TT (one rod with TT screws on each side, supplemented by a lateral satellite rod); and (c) 4R-TT/CBT (one rod with TT screws and one rod with CBT screws on each side). Loads of 400 N and 7.5 Nm in various directions were applied, with L5 fixed. Outcomes measured included range of motion (ROM) and maximum von Mises stress on rods and screws. RESULTS: The 4R-TT/CBT configuration had the smallest ROM in all directions, with reductions of 17.3%-26.4% compared to 2R-TT and 10.9%-16.8% compared to 4R-TT. It also showed the lowest von Mises stress on rods, particularly in flexion, where reductions ranged from 20.5% to 21.4% compared to 2R-TT and 20.9%-21.4% compared to 4R-TT. Considering the maximum von Mises stress on the screws, it exhibited the lowest and demonstrated a substantial reduction of 8.9%-22.8% and 19.5%-29.8% compared to the 2R-TT and 4R-TT configurations, respectively. CONCLUSION: The 4R-TT/CBT configuration provided superior biomechanical performance in lumbar PSO, enhancing construct stiffness, and substantially reducing stresses on rods and screws compared to the 2R-TT and 4R-TT configurations. However, this was a FE study with inherent simplifications. Therefore, the findings should be further confirmed by cadaveric or clinical studies in the future.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Biomechanical comparison of multirod constructs for severe thoracolumbar Pott's deformity following posterior vertebral column resection: a finite element study.Asian spine journal · 2026
- Cage-Orientation-Dependent Biomechanics of Endoscopic-Compatible Bidirectional Cortical Bone Trajectory-Modified Translaminar Facet Screw Fixation versus Bilateral Percutaneous Pedicle Screw Fixation in Osteoporotic TLIF: A Finite Element Study.Medical engineering & physics · 2026
- Biomechanical evaluation of various cage implantation angles and screw constructs in L4-5 PLIF surgery: a finite element study.BMC musculoskeletal disorders · 2026
- Biomechanical study of posterior lumbar fixation combined with partial laminectomy using finite element analysisBIO Web of Conferences · 2026
- Biomechanical comparison of laminoplasty and in-site regrafting versus laminectomy with pedicle screw fixation for localized thoracic pathological ossification of the ligamentum flavum: a finite element analysis.European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2026
- BIOMECHANICAL STUDY OF PULLOUT RESISTANCE OF CORTICAL SCREWS IN SWINE LUMBAR VERTEBRAE.Acta ortopedica brasileira · 2026