A Biomechanical Assessment of three Types of Plating Systems in a Combined Mandibular Parasymphysis and Angle Fracture - An In vitro Study.
Apoorva Basavaraj Kumbar, P Hema Mythily, Girish B Giraddi, K Sridevi
Department of Oral and Maxillofacial Surgery, Government Dental College and Research Institute, Bengaluru, Karnataka, India. · Department of Cleft and Craniofacial Surgery, Kanachur Hospital and Research Centre, Mangalore, Karnataka, India.
Key points
From the paper's conclusion- Plates which least resisted displacement between fractured segments were straight non-locking miniplates with a mean end-point force of 109.20N, as compared to 3D plates, 327N and locking plates, 345.2.
Abstract
INTRODUCTION: Mandibular fractures are the most encountered fractures by oral and maxillofacial surgeons, and for their fixation, different systems have been studied and used in clinical practice. Miniplates in lines of osteosynthesis described by Champy have stood the challenge of time for maxillofacial fractures. Many designs have been proposed since then. Three-dimensional (3D) plates, designed to have a geometric shape, claimed to add strength and stability to fracture fragments. Locking plates were designed to have screws that lock in the plate and thus provide added stability to fixation. MATERIALS AND METHODS: The aim of this study was to compare the efficacy of three internal fixation systems for osteosynthesis of combined angle and parasymphysis mandibular fractures. Fifteen sheep mandibles were used. Osteotomy simulating parasymphysis and contralateral angle fracture were done and fixed with three plating systems, 5 in each group. Group A with straight non-locking miniplates, Group B with 3D miniplates and Group C with locking miniplates. Universal Testing Machine was used to apply force ranging from 0 to 900 Newton (N) on the occlusal surface. The forces which caused significant displacement between fracture segments were recorded. The post hoc test showed significance between forces in the groups, i.e., P < 0.05. RESULTS: Plates which least resisted displacement between fractured segments were straight non-locking miniplates with a mean end-point force of 109.20N, as compared to 3D plates, 327N and locking plates, 345.2. DISCUSSION: Biomechanically, locking miniplates are more efficient than 3D miniplates which in turn are more efficient than non-locking miniplates in an in vitro scenario.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Influence of Plate Design, Thickness, and Fixation Architecture on Mandibular Advancement Stability: A Finite Element AnalysisJournal of Clinical Medicine · 2026
- Finite element analysis of patient specific three-dimensional titanium plate versus conventional two miniplate fixation for mandibular angle fractures.BMC oral health · 2026
- Biomechanical Comparison of Monoplanar and Biplanar Miniplate Fixation for Unfavorable Mandibular Angle Fractures: A Finite Element Analysis.Cureus · 2026
- A novel plate design versus two conventional miniplates for treatment of mandibular angle fractures (a pilot randomized controlled clinical trial)BMC Oral Health · 2026
- Evaluation of Locked Plates Versus 3-Dimensional Plates for Fixation of Mandibular Angle Fractures: A Clinical and Radiologic Study.The Journal of craniofacial surgery · 2026
- Does the biplane three-dimensional miniplate design provide stable fixation for mandibular angle fractures? A clinical study and proposal for novel radiographic criteria.Dental research journal · 2026