Early Complications of Epicardial Permanent Pacemaker Implantation in Neonates and Infants Weighing 5 kg or Less.
Eung Soo Kim, Youngkwan Song, Dong-Hee Kim, Eun Seok Choi, Bo Sang Kwon, Chun Soo Park, Tae-Jin Yun
Division of Pediatric Cardiac Surgery, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea. · Division of Pediatric Cardiac Surgery, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea. gninihs@amc.seoul.kr.
Key points
From the paper's conclusion- Early complications remain a significant concern after ePPM implantation in neonates and infants weighing 5kg or less, with body weight identified as an independent predictor. Patients weighing below 3.5 kg represent a high-risk subgroup requiring tailored surgical strategies and vigilant postoperative surveillance.
Abstract
BACKGROUND AND OBJECTIVES: We evaluated the incidence and predictors of early complications following epicardial permanent pacemaker (ePPM) implantation in neonates and infants weighing 5 kg or less. METHODS: We retrospectively reviewed 40 patients weighing 5 kg or less who underwent ePPM implantation between 1999 and 2025. Early complications were defined as ePPM-related adverse events requiring medical or surgical intervention within 6 months post-implantation. Firth's penalized logistic regression was performed to identify predictors for early complications. Receiver operating characteristic (ROC) determined the optimal cutoff value. RESULTS: The median age and body weight at implantation were 58 days (interquartile range [IQR], 21.8-118.3 days) and 3.7 kg (IQR, 3.2-4.3 kg), respectively. Early complications occurred in 9 patients (22.5%), including wound dehiscence (n=5), pocket infection (n=3), and pericardial effusion (n=1). On multivariable analysis, body weight at implantation was an independent predictor of complications (odds ratio [OR], 0.77 per 100-g increase; 95% confidence interval [CI], 0.60-0.98; p=0.031). ROC analysis identified 3.5 kg as the optimal cutoff value for predicting early complications (area under the curve, 0.94; 95% CI, 0.87-1.00; p<0.001), with 100% sensitivity and 84% specificity. A staged ePPM implantation strategy using initial temporary epicardial pacing was performed without early complications in four high-risk patients. CONCLUSIONS: Early complications remain a significant concern after ePPM implantation in neonates and infants weighing 5kg or less, with body weight identified as an independent predictor. Patients weighing below 3.5 kg represent a high-risk subgroup requiring tailored surgical strategies and vigilant postoperative surveillance.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Two Is Not too Small; a Case of a Ventricular Aveir AR2 for Ventricular Pacing in a 12.2-kg Two-Year-Old.Pediatric cardiology · 2026
- Lead-to-Vein Ratio and Long-Term Vascular Safety After Transvenous Pacing in Children Weighing ≤15 Kg.Pacing and clinical electrophysiology : PACE · 2026
- Risk factors for post-extubation stridor in pediatric cardiac intensive care unit: a single-center retrospective study.BMC pediatrics · 2026
- Repeated epicardial lead dislodgement in a child with trisomy 21 and complete atrioventricular block: A case report.Journal of cardiology cases · 2026
- Perioperative nursing care of a pediatric patient with 22q11.2 deletion syndrome (DiGeorge syndrome/velocardiofacial syndrome) after receiving leadless pacemaker implantation: a case reportFrontiers in Pediatrics · 2026
- Early Cardiopulmonary Physiology, Patent Ductus Arteriosus Characteristics, and Outcomes at the Limits of Viability.The Journal of pediatrics · 2026