Pediatric cardiology

Two Is Not too Small; a Case of a Ventricular Aveir AR2 for Ventricular Pacing in a 12.2-kg Two-Year-Old.

Nandita Sharma, Daniel Cortez

Department of Pediatrics, Davis School of Medicine, University of California, 2315 Stockton Blvd, Sacramento, CA, 95817, USA. · Department of Pediatrics, Davis School of Medicine, University of California, 2315 Stockton Blvd, Sacramento, CA, 95817, USA. dancortez@health.ucdavis.edu.

Case report

Key points

From the paper's conclusion
  • A 2-year-old, 12.2-kg female underwent successful right internal jugular vein implant of an Aveir AR2 into the right ventricle for backup ventricular pacing need.

Abstract

Pacemaker implants in pediatric patients can be filled with potential complications. Epicardial pacemaker implants are the usual mainstay for those under 15 kg and transvenous devices can have infectious and pocket erosion risk. Retrievable leadless pacemakers seem to be a potential option. Retrospective review of a 2-year-old patient with high grade AV block and overall lowering average junctional rate was performed. Internal review board approval was obtained. A 2-year-old, 12.2-kg female underwent successful right internal jugular vein implant of an Aveir AR2 into the right ventricle for backup ventricular pacing need. No complications occurred. Follow-up at 12 months demonstrated a threshold of 0.5 Volts @0.2 ms, impedance at 420 ohms and sensing at 10milliVolts with 30% pacing noted with programming at VVI 50 bpm, with predicted longevity at 20.2 years. Feasibility of internal jugular vein implant of an Aveir AR2 is possible without complication in a 2-year-old, 12.2 kg child. Further prospective studies are needed to assess true feasibility in this population.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers