Two Is Not too Small; a Case of a Ventricular Aveir AR2 for Ventricular Pacing in a 12.2-kg Two-Year-Old.
Nandita Sharma, Daniel Cortez
Department of Pediatrics, Davis School of Medicine, University of California, 2315 Stockton Blvd, Sacramento, CA, 95817, USA. · Department of Pediatrics, Davis School of Medicine, University of California, 2315 Stockton Blvd, Sacramento, CA, 95817, USA. dancortez@health.ucdavis.edu.
Key points
From the paper's conclusion- A 2-year-old, 12.2-kg female underwent successful right internal jugular vein implant of an Aveir AR2 into the right ventricle for backup ventricular pacing need.
Abstract
Pacemaker implants in pediatric patients can be filled with potential complications. Epicardial pacemaker implants are the usual mainstay for those under 15 kg and transvenous devices can have infectious and pocket erosion risk. Retrievable leadless pacemakers seem to be a potential option. Retrospective review of a 2-year-old patient with high grade AV block and overall lowering average junctional rate was performed. Internal review board approval was obtained. A 2-year-old, 12.2-kg female underwent successful right internal jugular vein implant of an Aveir AR2 into the right ventricle for backup ventricular pacing need. No complications occurred. Follow-up at 12 months demonstrated a threshold of 0.5 Volts @0.2 ms, impedance at 420 ohms and sensing at 10milliVolts with 30% pacing noted with programming at VVI 50 bpm, with predicted longevity at 20.2 years. Feasibility of internal jugular vein implant of an Aveir AR2 is possible without complication in a 2-year-old, 12.2 kg child. Further prospective studies are needed to assess true feasibility in this population.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Aveir VR for Atrial Leadless Pacing in Young Patients.Pediatric cardiology · 2026
- Jugular Journeys: A Novel Approach for AVEIR Leadless Pacemaker ImplantationPacing and Clinical Electrophysiology · 2026
- Contemporary Advances of Leadless Pacing and Conduction System Pacing in Children With Congenital Heart Disease.Pacing and clinical electrophysiology : PACE · 2026
- Successful atrial leadless pacemaker implantation in an 11-year-old with congenital sinus node dysfunction: a case report.European heart journal. Case reports · 2026
- Lead-to-Vein Ratio and Long-Term Vascular Safety After Transvenous Pacing in Children Weighing ≤15 Kg.Pacing and clinical electrophysiology : PACE · 2026
- Early Complications of Epicardial Permanent Pacemaker Implantation in Neonates and Infants Weighing 5 kg or Less.Korean circulation journal · 2026