Lead-to-Vein Ratio and Long-Term Vascular Safety After Transvenous Pacing in Children Weighing ≤15 Kg.
Michel Cabrera Ortega, Juan Carlos Ramiro Novoa, Osmin Castañeda Chirino, Adel Eladio González Morejón, Giselle Serrano Ricardo, Yoel Coto Hernández, Mabel Elena Dominguez Gonzalez, Liz Coto Pérez,
Department of Cardiology, Parc Sanitari Sant Joan de Deu, Barcelona, Spain. · Cardiac Catheterization Laboratory, Cardiocentro Pediatrico William Soler, La Habana, Cuba.
Key points
From the paper's conclusion- During the first transvenous system lifespan, imaging-defined venous abnormalities were relatively frequent but mostly asymptomatic, whereas clinically overt vascular complications were uncommon. The lead-to-vein ratio may support anatomical pre-implant risk stratification, although multicenter validation and follow-up beyond generator replacement are required.
Abstract
BACKGROUND: Transvenous pacing in low-weight children is technically feasible, but its clinical use depends on long-term venous patency, system performance, and ventricular function. We aimed to describe imaging-defined venous abnormalities and clinically overt vascular events after first transvenous pacemaker implantation and to explore whether the lead-to-vein ratio provides anatomical information beyond body weight. METHODS: This prospective observational cohort included 35 children weighing ≤15 kg undergoing first transvenous pacemaker implantation between 2010 and 2020. The primary vascular endpoint was imaging-defined venous thrombosis, occlusion, or stenosis ≥50%; clinical impact was recorded separately. Device-related safety, minimum venous diameter, lead-to-vein ratio, ventricular function, synchrony, and paced QRS duration were analyzed using non-parametric, exact, and exploratory univariable tests. RESULTS: Median follow-up was 6.7 [4.5-8.2] years. Eleven patients weighed <10 kg and 3 weighed <5 kg. The imaging-defined vascular endpoint occurred in 6/35 patients; 5 were asymptomatic surveillance findings and 1 was clinically symptomatic. The lead-to-vein ratio was higher in patients with the endpoint than in those without events (0.59 [0.55-0.63] vs. 0.47 [0.43-0.51]; p = 0.003). A ratio >0.55 was associated with more frequent imaging-defined endpoints (4/8 vs. 2/27; odds ratio 12.5; 95% confidence interval 1.7-92.3). Left ventricular ejection fraction remained preserved. CONCLUSION: During the first transvenous system lifespan, imaging-defined venous abnormalities were relatively frequent but mostly asymptomatic, whereas clinically overt vascular complications were uncommon. The lead-to-vein ratio may support anatomical pre-implant risk stratification, although multicenter validation and follow-up beyond generator replacement are required.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Two Is Not too Small; a Case of a Ventricular Aveir AR2 for Ventricular Pacing in a 12.2-kg Two-Year-Old.Pediatric cardiology · 2026
- Early Complications of Epicardial Permanent Pacemaker Implantation in Neonates and Infants Weighing 5 kg or Less.Korean circulation journal · 2026
- Contemporary Advances of Leadless Pacing and Conduction System Pacing in Children With Congenital Heart Disease.Pacing and clinical electrophysiology : PACE · 2026
- Leadless Pacemaker Implant Location as a Predictor of Pacing Induced Cardiomyopathy.Pacing and clinical electrophysiology : PACE · 2026
- Multicenter study on electrical performance and predictive indicators in aveir VR pacemaker recipients.Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing · 2026
- Leadless Pacemaker, Right Ventricular Pacing Burden, and Heart Failure.Heart rhythm · 2026