Ultrasonographic Pupil Assessment in Patients With Severe Traumatic Brain Injury.

Qian Shi, Rujing Bai, Yan Xu, Bo Feng, Ling Yuan, Niannian Liu, Danjiang Dong

Qian Shi is a staff nurse, Intensive Care Unit, Nanjing Drum Tower Hospital, Nanjing, China. · Rujing Bai is a ward nurse manager, Intensive Care Unit, Nanjing Drum Tower Hospital, Nanjing, China.

Observational

Key points

From the paper's conclusion
  • Ultrasonographic pupil assessment is a reliable and accurate method for objective pupillary evaluation when direct visualization is obstructed. By enabling quantitative measurement of the pupillary light reflex through the closed eyelid, ultrasonographic pupil assessment addresses an important monitoring gap in the management of patients with severe traumatic brain injury and periorbital swelling.

Abstract

BACKGROUND: In patients with severe traumatic brain injury, accurate pupillary assessment is critical for neurological monitoring but is often hindered by periorbital edema, trauma, or dressings that preclude direct visualization. OBJECTIVE: To validate the feasibility and accuracy of ultrasonographic pupil assessment performed through the closed eyelid as an alternative to direct pupillometry in patients with traumatic brain injury. METHODS: This prospective study enrolled 45 intensive care unit patients (36 male; mean [SD] age, 67.1 [12.9] years). Both direct pupillometry and ultrasonographic pupil assessment were sequentially completed within a 3-minute interval under standardized low-light conditions for both eyes. Direct pupillometry involved video recording of the exposed pupil with a calibrated reference scale. Ultrasonographic pupil assessment was performed by using a high-frequency linear probe placed on the closed eyelid. The primary outcome was the agreement between methods for calculating the percentage of pupillary constriction. Secondary outcomes included agreement on resting and minimum pupil diameters. RESULTS: Ultrasonographic pupil assessment showed excellent agreement with direct pupillometry for percent pupillary constriction (median [IQR] difference, 0.44% [-2.78% to 2.83%]; Wilcoxon P ≥ .48). Bland-Altman analysis showed narrow limits of agreement. Strong correlations were also noted for resting and minimum pupil diameters (0.6 ≤ r2 ≤ 0.873). CONCLUSION: Ultrasonographic pupil assessment is a reliable and accurate method for objective pupillary evaluation when direct visualization is obstructed. By enabling quantitative measurement of the pupillary light reflex through the closed eyelid, ultrasonographic pupil assessment addresses an important monitoring gap in the management of patients with severe traumatic brain injury and periorbital swelling.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers