In-the-Bag Stability of a New Hydrophilic Acrylic Toric Intraocular Lens.
Nikolaus Mahnert, Marcus Lisy, Markus Schranz, Victor Danzinger, Claudette Abela-Formanek, Christina Leydolt, Rupert Menapace, Daniel Schartmüller
Department of Ophthalmology and Optometry, Medical University of Vienna, Vienna, Austria.
Key points
From the paper's conclusion- Although mean absolute tilt and decentration values were low, in some cases, a pronounced tilt and posterior axial hyperopic shift occurred. Excessive anterior fibrosis development was responsible for these changes.
Abstract
PURPOSE: To assess tilt, decentration, and axial stability of the RayOne EMV toric intraocular lens (TIOL) (Rayner Intraocular Lenses Ltd) over a 6-month follow-up period. METHODS: This was a single-center, prospective, interventional clinical trial at the Department of Ophthalmology and Optometry, Medical University of Vienna, Vienna, Austria. A total of 119 eyes from 79 patients with age-related cataract were implanted with a RayOne EMV TIOL. Tilt, decentration, and aqueous depth (AQD) were measured using anterior segment optical coherence tomography (Casia 2; Tomey Corporation) preoperatively, at 1 hour, 1 week, and 6 months postoperatively. Individual changes in tilt and decentration were evaluated using vector analysis. RESULTS: Mean absolute tilt and decentration values were 5.30 ± 1.38° and 0.14 ± 0.10 mm for the crystalline lens, 5.28 ± 1.27° and 0.21 ± 0.11 mm for the TIOL at 1 hour, 5.22 ± 1.25° and 0.20 ± 0.10 mm at 1 week, 5.33° ± 1.81° (range: 0.90 to 11.10°) and 0.22 ± 0.11 mm (range: 0.04 to 0.57 mm) at 6 months. Mean changes in tilt and decentration from preoperatively to 6 months were 1.59 ± 1.66° (range: 0.09 to 8.43°) and 0.17 ± 0.09 mm (range: 0.02 to 0.47 mm). Seven TIOLs showed a tilt change over 5° within 6 months. A posterior axial shift of 0.13 ± 0.14 mm (range: -0.06 to 0.69 mm) was observed between 1 week and 6 months and correlated with anterior fibrosis severity (P < .001). CONCLUSIONS: Although mean absolute tilt and decentration values were low, in some cases, a pronounced tilt and posterior axial hyperopic shift occurred. Excessive anterior fibrosis development was responsible for these changes.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Rotational stability and visual outcomes of toric implantable collamer lenses in eyes with horizontal and non-horizontal implanting orientation: an observational study.International ophthalmology · 2026
- Early refractive stability and predictive accuracy of a novel hydrophobic C-loop monofocal intraocular lens in eyes with normal axial length.Medicine · 2026
- Influencing Factors of Positional Stability of Implantable Collamer Lenses in Terms of Rotation, Decentration, and Tilt in Patients With Myopia.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- Long-term visual outcomes and rotational stability of two diffractive toric intraocular lenses with different design combinations in cataract patients with long axial length.Frontiers in medicine · 2026
- Longitudinal agreement between retroillumination-based and scleral vessel landmark-based measurement of toric intraocular lens rotational drift using CASIA2: a prospective within-subject comparison study.BMC ophthalmology · 2026
- Clinical Performance of an Isofocal Toric Intraocular Lens with Double C-Loop Haptics: A Multicenter Retrospective Study.Clinical ophthalmology (Auckland, N.Z.) · 2026