Clinical Performance of an Isofocal Toric Intraocular Lens with Double C-Loop Haptics: A Multicenter Retrospective Study.
Sheraz Daya, Takayuki Akahoshi, Camille Bosc, Alessandra Di Maria, Marcela Espinosa Lagana, Noemi Sasso, Alessandro Gaeta
Centre for Sight, East Grinstead, UK. · Nihonbashi Cataract-Clinic, Tokyo, Japan.
Key points
From the paper's conclusion- Eyes with pre-existing astigmatism implanted with the Isopure Serenity toric IOL showed good distance visual acuity and functional intermediate vision, with high refractive accuracy and good rotational stability following cataract and refractive lens exchange surgery.
Abstract
PURPOSE: To evaluate clinical outcomes in eyes with pre-existing astigmatism undergoing cataract surgery or refractive lens exchange with the implantation of an isofocal optic-design toric intraocular lens (IOL) with double C-loop haptics. METHOD: This multicenter retrospective study comprised 100 eyes (65 patients) implanted with the Isopure Serenity toric IOL (BVI, Inc.) conducted across four sites. Postoperative evaluation at between 2 and 3 months included manifest refraction, monocular uncorrected and distance corrected visual acuity (UDVA and CDVA), monocular uncorrected and distance corrected intermediate visual acuity (UIVA and DCIVA, at 80 and 66 cm), monocular uncorrected and distance corrected near visual acuity (UNVA and DCNVA, at 40 cm), binocular defocus curve and rotational stability. RESULTS: Manifest refraction spherical equivalent and refractive astigmatism were -0.24±0.41D and -0.36±0.36D, respectively. About 79% and 97% of eyes were within ±0.50D and ±1.00D, respectively, and 80% and 96% of eyes were ≤0.50D and ≤1.00D of refractive cylinder, respectively. At distance vision, the mean monocular CDVA was -0.02±0.06 logMAR and 92.93% of eyes showed a CDVA of ≥20/20. At intermediate vision, the mean monocular DCIVA values were 0.10±0.07 and 0.17±0.10 logMAR at 80 cm and 66 cm, respectively; and 96.92% and 88.33% of eyes showed a DCIVA of ≥20/32 at 80 cm and 66 cm, respectively. At near vision, the mean monocular DCNVA was 0.31±0.08 logMAR and 75.44% of eyes showed a DCNVA of ≥20/40. The binocular depth of focus range was approximately 1.50D. The mean absolute IOL rotation was 2.73±2.58 degrees with all eyes having a rotation of <10 degrees. CONCLUSIONS: Eyes with pre-existing astigmatism implanted with the Isopure Serenity toric IOL showed good distance visual acuity and functional intermediate vision, with high refractive accuracy and good rotational stability following cataract and refractive lens exchange surgery.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Use of the internal astigmatism axis to determine the toric intraocular lens axis: Comparison between C-loop and plate-haptic lenses.Arquivos brasileiros de oftalmologia · 2026
- Early Visual Outcomes of Toric Versus Non-Toric IOLs in Low-to-Moderate Corneal Astigmatism: A Real-World Matched Cohort Study.Journal of clinical medicine · 2026
- Visual and refractive outcomes of eyes with low corneal astigmatism implanted with low toric or non-toric trifocal intraocular lensesBMC Ophthalmology · 2026
- Bilateral Refractive Lens Exchange with a Hydrophobic Trifocal Intraocular Lens.Clinical ophthalmology (Auckland, N.Z.) · 2026
- Influence of Astigmatism Magnitude and Axis on the Selection of Non-Toric versus Low-Powered Trifocal Toric Intraocular Lenses in Presbyopic Refractive Lens Exchange.Journal of cataract and refractive surgery · 2026
- Visual and refractive outcomes of Eyecryl phakic intraocular lens in myopia and myopic astigmatism.Indian journal of ophthalmology · 2026