Configuration-Specific Temporal Alignment of a Repurposed Wearable Fuel-Cell Alcohol Sensor with an MQ-3 Timing Proxy in a Laboratory Static-Headspace Chamber.
Hasan Mirzakhaninafchi, Glen Rains, Wesley Porter, Canicius Mwitta, Denis Kiobia, Shekhar Thapa, Peter Ngimbwa, Christopher Ranger
Department of Entomology, University of Georgia, Tifton, GA 31793, USA. · Department of Crop and Soil Sciences, University of Georgia, Tifton, GA 31793, USA.
Key points
From the paper's conclusion- We selected the BACtrack Skyn wearable fuel-cell alcohol sensor for its compact form and timestamped output as a candidate for future tree-headspace measurements, and evaluated one device with an MQ-3 metal-oxide timing proxy and environmental logging in a closed static-headspace chamber.
Abstract
Ethanol emitted from stressed woody tissues is a host-associated cue for ambrosia beetles, motivating investigation of portable ethanol-responsive sensors for future tree monitoring. This laboratory study evaluated temporal alignment between two asynchronously logged volatile organic compound (VOC)-responsive sensors during a denatured-alcohol chamber challenge. We selected the BACtrack Skyn wearable fuel-cell alcohol sensor for its compact form and timestamped output as a candidate for future tree-headspace measurements, and evaluated one device with an MQ-3 metal-oxide timing proxy and environmental logging in a closed static-headspace chamber. The dataset comprised five alcohol-free control runs (B0-B4) and three technical 1 μL denatured-alcohol challenge runs (EtOH1-EtOH3). Signals were processed on a common 1 min grid using a 10-15 min analytical baseline and a 15-45 min response-characterization window. MQ-3 peaks occurred at 17.0-17.3 min from the run origin (T0), whereas Skyn device-output peaks occurred at 18.8-19.2 min; peak-time offsets were 1.5-2.2 min. First-difference analysis selected a 2 min alignment setting on the primary 1 min grid in all three challenge runs; a supplementary 20 s analysis gave run-specific optima of 80-160 s. Across the three challenge runs, run-specific Pearson correlations ranged from 0.686 to 0.751 before alignment and from 0.861 to 0.930 after alignment; each coefficient was calculated from the paired 1 min observations in the 15-45 min response window. Normalized root-mean-square error (RMSE) ranged from 0.208 to 0.272 before alignment and from 0.121 to 0.230 after alignment, whereas rank correlation did not improve consistently. Alcohol-free controls remained below challenge-run peak magnitudes, although MQ-3 traces included drift and positive excursions. These within-run comparisons describe alignment in the tested configuration, not intrinsic sensor latency, concentration calibration, or validated tree-monitoring performance.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Compensation of Ambient Variations in a Temperature-Modulated SnO2 Gas Sensor-Based Electronic Nose for Accurate Discrimination of Structurally Similar Volatile Organic Compounds.ACS sensors · 2026
- Field Evaluation of a Portable Multi-Sensor Platform for Characterizing Inter-Individual Variability in Exhaled Breath Responses.Sensors (Basel, Switzerland) · 2026
- Development and LoRa-based implementation of a ZIF-8 chemiresistive CO2 sensor for air quality monitoring.RSC advances · 2026
- Removal of Ethanol from Indoor Air by Ficus elastica Roxb.: Process Optimisation and Post-Removal Desorption Dynamics.Plants (Basel, Switzerland) · 2026
- Retention of Tritiated Alcohols by Zeolites Used in Indoor Air Sampling.ACS omega · 2026
- B-395 Beyond Ethanol: Leveraging Routine Alcohol Testing to Detect Difluoroethane and Enhance Patient SafetyClinical Chemistry · 2026