Design of a Low-Cost Auto-Ranging Electrometer for Current-Mode Dosimetry.
Marko Andjelković, Goran S Ristić
IHP-Leibniz Institute for High Performance Microelectronics, Im Technologiepark 25, 15236 Frankfurt (Oder), Germany. · Faculty of Electronic Engineering, University of Niš, Aleksandra Medvedeva 4, 18000 Niš, Serbia.
Key points
From the paper's conclusion- The results have shown that the realized electrometer, combined with the investigated PIN photodiodes, can serve as a low-cost current-mode dosimeter for measuring the gamma radiation dose rates from 0.9 Gy/h to 34 Gy/h, and absorbed doses at least up to 33 Gy.
Abstract
This paper presents a low-cost auto-ranging electrometer for current-mode dosimetric applications. The proposed electrometer is based on a variable gain transimpedance amplifier, and has been realized with commercially available components. It can operate as a stand-alone unit with an integrated display for real-time measurement visualization, or can be interfaced to a personal computer (PC) over USB or RS-232 connections, allowing real-time visualization and data storage using custom-made PC application software. The electrometer is capable of measuring direct currents from 50 pA to 10 µA in three sensitivity ranges, with relative error less than 2.5%, and response time of approximately 0.2 s in each range. The performance of the proposed electrometer is comparable with custom-made electrometers reported in the literature, as well as with several standard commercial dosimetric electrometers. The electrometer has been validated experimentally for measuring the direct current of four commercial PIN photodiodes under gamma irradiation from a Co-60 source. The results have shown that the realized electrometer, combined with the investigated PIN photodiodes, can serve as a low-cost current-mode dosimeter for measuring the gamma radiation dose rates from 0.9 Gy/h to 34 Gy/h, and absorbed doses at least up to 33 Gy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Feasibility study of in‐house second‐channel calibration of dual‐channel electrometers using a battery‐powered current sourceJournal of Applied Clinical Medical Physics · 2026
- Characteristics of electrometers on ionization chamber dosimetry in ultra-high dose-rate proton beam irradiation.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Development of a real-time CMOS gamma radiation dosimeter sensor.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Long-term stability evaluation of user-derived electrometer calibration factors (kelec, user) for radiotherapy dosimetry.Journal of applied clinical medical physics · 2026
- Impact of a common user-derived electrometer calibration coefficient on absorbed dose-to-water determination: a multicenter uncertainty studyJournal of Radioanalytical and Nuclear Chemistry · 2026
- Evaluation of the Exradin A30 Parallel Plate Ion Chamber as a Reference Dosimeter in Ultra-High Dose Rate (UHDR) Electron Beams.ArXiv · 2026