Characteristics of electrometers on ionization chamber dosimetry in ultra-high dose-rate proton beam irradiation.
Ryosaku Yamada, Teiji Nishio
Department of Radiology, Kouseikai Takai Hospital, 470-8, Kuranosho, Tenri-shi, Nara 632-0006, Japan. · Medical Physics Laboratory, Division of Health Science, Graduate School of Medicine, The University of Osaka, 1-1, Yamadagaoka, Suita-shi, Osaka 565-0871, Japan. Electronic address: nishio@sahs.med.osaka-u.ac.jp.
Key points
From the paper's conclusion- Our findings suggest that current integration-type electrometers perform better with conventional dose rates, whereas charge accumulation-type electrometers allow for more stable measurements in UHDR environments.
Abstract
PURPOSE: This study investigated the influence of differences in electrometers on the dosimetry of UHDR scanning proton beam irradiation. METHOD: A measurement system that combined ionization measurements using an ionization chamber and electrometers with luminescence observations of an intensifying screen was used. Proton beams were delivered to this system at various doses and dose rates, and dose-to-luminescence ratios were calculated. Performance was compared among three different electrometers: RAMTEC DUO (RAMTEC), Sakura Proof (Sakura), which are charge accumulation-type, and UNIDOS webline (UNIDOS), which is current integration-type. RESULTS: The differences between the maximum and minimum dose-to-luminescence values were 0.851 % (4 Gy) and 1.543 % (20 Gy) for the RAMTEC, 0.913 % (4 Gy) and 1.626 % (20 Gy) for the Sakura, and 1.885 % (4 Gy) and 1.672 % (20 Gy) for the UNIDOS. A comparison of the variances showed that the UNIDOS exhibited significantly larger variances at 20.0, 50.0, and 100.0 Gy/s for 4 Gy and at 124.0 and 167.0 Gy/s for 20 Gy. Regarding dose rate correlations, at 4 Gy, RAMTEC and Sakura showed a positive correlation at dose rates of 10 Gy/s or less. At 20 Gy, no difference was observed among the three electrometers for dose rates up to 50 Gy/s. However, only the UNIDOS showed a negative correlation at dose rates of 100 Gy/s or higher. CONCLUSION: Our findings suggest that current integration-type electrometers perform better with conventional dose rates, whereas charge accumulation-type electrometers allow for more stable measurements in UHDR environments.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of the Exradin A30 Parallel Plate Ion Chamber as a Reference Dosimeter in Ultra-High Dose Rate (UHDR) Electron Beams.ArXiv · 2026
- Long-term stability evaluation of user-derived electrometer calibration factors (kelec, user) for radiotherapy dosimetry.Journal of applied clinical medical physics · 2026
- Design of a Low-Cost Auto-Ranging Electrometer for Current-Mode Dosimetry.Sensors (Basel, Switzerland) · 2026
- Dose-rate metrics for ionization-chamber recombination in synchrotron-based ultra-high-dose-rate proton and carbon-ion beams: application and validity of quasi-static response models.Zeitschrift fur medizinische Physik · 2026
- Feasibility study of in‐house second‐channel calibration of dual‐channel electrometers using a battery‐powered current sourceJournal of Applied Clinical Medical Physics · 2026
- Impact of a common user-derived electrometer calibration coefficient on absorbed dose-to-water determination: a multicenter uncertainty studyJournal of Radioanalytical and Nuclear Chemistry · 2026