Evaluation of the Exradin A30 Parallel Plate Ion Chamber as a Reference Dosimeter in Ultra-High Dose Rate (UHDR) Electron Beams.
Mervat Alharbi, Kevin Liu, Brian Hooten, Shannon Holmes, Stefanno Alarcon-Nunez, Jeffrey Radtke, Larry DeWerd, Sam Beddar,
Key points
From the paper's conclusion- This study evaluated the commercial A30 as a reference dosimeter for UHDR electron beams by characterizing leakage current, CCE, polarity correction (Ppol), and beam-quality correction factors (kQ).
Abstract
Reliable reference dosimetry for ultra-high dose-rate (UHDR) beams (>40 Gy/s) is challenging because conventional ionization chambers (ICs) exhibit saturation from ion recombination. The Exradin A30 IC uses an ultra-thin 0.3-mm electrode spacing to improve charge-collection efficiency (CCE). This study evaluated the commercial A30 as a reference dosimeter for UHDR electron beams by characterizing leakage current, CCE, polarity correction (Ppol), and beam-quality correction factors (kQ). Measurements were performed with a 9-MeV IntraOp Mobetron from the accelerator head, achieving up to 9 Gy per pulse (DPP) and an instantaneous dose rate of 2.25 MGy/s. Data were acquired in grounded water-equivalent plastic, distilled water, and saline water. DPP was varied by changing SSD at a fixed 4-μs pulse width, while pulse repetition frequency (PRF) ranged from 5 to 90 Hz. CCE was determined using EBT-XD film under matched UHDR and conventional dose and energy conditions. CCE and Ppol were also evaluated as functions of DPP and PRF in distilled and saline water. Values of kQ were calculated using Monte Carlo simulations and measured in TrueBeam electron beams. Leakage current was <2 fA. Both CCE and Ppol decreased with increasing DPP; however, CCE remained 90-99% across all three phantoms, while Ppol decreased from 0.990 to 0.981 in liquid and solid water. Neither CCE nor Ppol depended on PRF over 5-90 Hz. Measured and calculated kQ values agreed within 0.8% at all energies except 9 MeV, where they differed by 2%. The A30 exhibited 5% recombination at DPP up to 5 Gy in distilled and saline water. Its response in solid phantoms was affected by charge buildup, which was mitigated by grounding. With appropriate CCE corrections and grounded solid phantoms, the commercial A30 is suitable for reference dosimetry in UHDR electron beams.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Dose-rate metrics for ionization-chamber recombination in synchrotron-based ultra-high-dose-rate proton and carbon-ion beams: application and validity of quasi-static response models.Zeitschrift fur medizinische Physik · 2026
- Characteristics of electrometers on ionization chamber dosimetry in ultra-high dose-rate proton beam irradiation.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Charge collection efficiency in air-vented plane-parallel ionisation chambers at ultra-high dose rates: a self-consistent Garfield++ Monte Carlo model including space-charge effects and ion recombinationPhysics in Medicine and Biology · 2026
- Impact of a common user-derived electrometer calibration coefficient on absorbed dose-to-water determination: a multicenter uncertainty studyJournal of Radioanalytical and Nuclear Chemistry · 2026
- Study of three-dimensional dose distribution measurement of ultra-high dose rate electron beam.Physics in medicine and biology · 2026
- Toward quantitative reconstruction of delivered dose in a 3D digital water phantom for electron FLASH-RT via annular-array electron-induced acoustic imaging.Physics in medicine and biology · 2026