Multicenter evaluation of an innovative fully-automated molecular diagnostics system.
Yaping Ren, Chao Chen, Weizhong Wang, Zhengquan Yang, Zhonggang Fang, Ying Yu, Zhongyuan Xiang
Department of Laboratory Medicine, The Second Xiangya Hospital of Central South University, Changsha, China. · Institute of In Vitro Diagnostics Technology at Snibe Diagnostic, Shenzhen New Industries Biomedical Engineering (Snibe) Co. Ltd. Shenzhen, China.
Key points
From the paper's conclusion- Excellent analytical characteristics were verified in various analyte and sample types for Molecision® R8. Clinical safety and effectiveness of R8 were substantiated by its strong consistency with other polymerase chain reaction (PCR) platforms in routine clinical use. With sample-in-result-out workflow, Molecision® R8 could further improve the efficacy of clinical molecular laboratories.
Abstract
OBJECTIVES: To evaluate the analytical performance of a novel fully-automated molecular diagnostics system Molecision® R8, together with its multiplex testing capability and compatibility with different types of analytes and samples. METHODS: Precision, limit of detection (LoD) and linearity were evaluated based on respective guidelines. Method comparisons were conducted between Molecision® R8 and two commercially available systems Sansure and Hybribio, using four assays for use with each system - hepatitis B virus (HBV) RNA, Chlamydia trachomatis (CT) DNA, Ureaplasma urealyticum (UU) DNA, and Neisseria gonorrhoeae (NG) DNA. RESULTS: Molecision® R8 demonstrated excellent analytical performance: within-lab coefficient of variation (CV) was ≤1.55 %, with 100 % detectable rates at claimed LoD and strong linearity with correlation coefficient of 0.9998. In HBV RNA comparison between Molecision® R8 and Sansure system, the 95 % confidence interval of the slope and the intercept in Passing-Bablok regression contained 1 and 0 respectively, and mean bias was -0.05 lg IU/mL in Bland-Altman analysis. Over 95 % agreement rates were found between R8 and Hybribio system in CT DNA, UU DNA and NG DNA detection, except for the negative agreement rate of UU DNA. CONCLUSIONS: Excellent analytical characteristics were verified in various analyte and sample types for Molecision® R8. Clinical safety and effectiveness of R8 were substantiated by its strong consistency with other polymerase chain reaction (PCR) platforms in routine clinical use. With sample-in-result-out workflow, Molecision® R8 could further improve the efficacy of clinical molecular laboratories.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comprehensive Performance Verification of the Mindray MN2880 System in Detecting Respiratory Virus Nucleic Acids.Journal of clinical laboratory analysis · 2026
- Performance Evaluation of HBV RNA Real-Time PCR Kits for Quantitative Detection and Clinical Comparison Across Different Detection SystemsMicroorganisms · 2026
- Performance evaluation and application of a multiplex PCR capillary electrophoresis method for detecting nucleic acids of seven sexually transmitted pathogensFrontiers in Cellular and Infection Microbiology · 2026
- Concordance Between Multiplex-PCR Resistance-Gene Detection and MIC-Based Susceptibility in Gram-Negative Clinical Isolates from a Romanian Tertiary Hospital.Antibiotics (Basel, Switzerland) · 2026
- Diagnostic Performance of the FilmArray BCID2 Multiplex PCR Panel Applied to Excised Heart Valve Specimens from Patients with Infective Endocarditis.Microorganisms · 2026
- Comparative performance evaluation of two rapid molecular assays for detecting drug-resistant tuberculosis in clinical isolates.Microbiology spectrum · 2026