Can Autorefractometer and Digital Pupillometer Readings of Interpupillary Distance Be Used Interchangeably?: A Cross-Sectional Study.
Mansour A Alghamdi
Department of Optometry, College of Applied Medical Sciences, King Saud University, Riyadh 11451, Saudi Arabia.
Key points
From the paper's conclusion- This cross-sectional study evaluated the agreement between IPD measurements obtained using an autorefractometer and a handheld digital pupillometer.
Abstract
Accurate measurement of interpupillary distance (IPD) is critical for spectacle prescription and binocular vision assessment. This cross-sectional study evaluated the agreement between IPD measurements obtained using an autorefractometer and a handheld digital pupillometer. A total of 88 participants (aged 8-64 years, including 14 aged <18 years) underwent IPD assessment with both instruments under standardized conditions. Agreement was assessed using Bland-Altman analysis with 95% confidence intervals (CIs) for the limits of agreement (LOA), the intraclass correlation coefficient (ICC) for absolute agreement, and a clinical tolerance of ±2.5 mm based on ANSI Z80.1-2020. The mean difference between devices was small and not statistically significant (0.27 mm; 95% CI -0.09 to 0.64 mm; p = 0.14), and the ICC was 0.84 (95% CI 0.77-0.89). However, the 95% LOA were wide (-3.10 to +3.64 mm) and exceeded the clinical tolerance, and 12.5% of participants had between-device differences greater than 2.5 mm. The differences were not associated with age or sex. Although both devices agree well on average, they should not be considered interchangeable for individual patients, particularly where high-precision fitting is required, such as for multifocal lenses and virtual reality systems. Consistent use of the same device across patient encounters is recommended.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A comparative analysis of pupillometry measurements obtained from different devices featuring static and dynamic capabilities.Documenta ophthalmologica. Advances in ophthalmology · 2026
- Inter-device Agreement of Device-Designated Chord Mu and Chord Alpha Measurements in Patients Undergoing Preoperative Cataract Evaluation.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- Agreement between digital ocular alignment measurement devices and the prism alternate cover test in horizontal strabismus.International ophthalmology · 2026
- Pupil size measurements with two multifunctional devices based on different technologies and an infrared based pupillometer.Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie · 2026
- Variability in Refractive Measurements in Children with Autism Spectrum Disorder Under Cycloplegic and Non-Cycloplegic Conditions: Influence of Measurement Method.Journal of clinical medicine · 2026
- Effect of fogging lenses on accommodation using an open-view portable wavefront autorefractor.Journal of optometry · 2026