Agreement between digital ocular alignment measurement devices and the prism alternate cover test in horizontal strabismus.
Orkun Özkan, Halil Hüseyin Çağatay, Bağdagül Koçak
Department of Ophthalmology, Faculty of Medicine, Uskudar University, Istanbul, Turkey. orkunozkan33@gmail.com. · Private Practice, Izmir, Turkey.
Key points
From the paper's conclusion- Digital ocular alignment devices demonstrated moderate agreement with PACT, particularly in patients with moderate-angle deviations. However, agreement decreased in larger deviations, and the wide limits of agreement indicate that individual measurements should be interpreted with caution. Therefore, PACT remains the clinical reference method for accurate deviation measurement, particularly in surgical planning.
Abstract
PURPOSE: To evaluate the agreement between digital ocular alignment devices and the prism alternate cover test (PACT) in patients with horizontal strabismus. METHODS: This retrospective cross-sectional study included 63 patients (31 esotropia, 32 exotropia). Ocular deviation was measured using the prism alternate cover test (PACT) at near (CTN) and distance (CTD) and compared with measurements obtained using the Welch Allyn Spot Vision Screener and the BCN Innova Sensor. Agreement was assessed using intraclass correlation coefficients and Bland-Altman analysis. Subgroup analyses were performed according to deviation magnitude (10-30 PD and >30 PD). RESULTS: The estimated ICC values indicated moderate agreement between the digital devices and PACT measurements. Agreement was generally stronger in exotropia than in esotropia. Bland-Altman analysis demonstrated small mean biases but wide limits of agreement. Agreement was more consistent within the 10-30 PD range and decreased in larger deviations. CONCLUSIONS: Digital ocular alignment devices demonstrated moderate agreement with PACT, particularly in patients with moderate-angle deviations. However, agreement decreased in larger deviations, and the wide limits of agreement indicate that individual measurements should be interpreted with caution. Therefore, PACT remains the clinical reference method for accurate deviation measurement, particularly in surgical planning.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Smartphone-Based Digital Ruler to Automatically Measure Strabismus in Ophthalmologist-Level: A Prospective, Multicenter Cohort Study.NEJM AI · 2025
- Can Autorefractometer and Digital Pupillometer Readings of Interpupillary Distance Be Used Interchangeably?: A Cross-Sectional Study.Life (Basel, Switzerland) · 2026
- Squint ASSIST: A Novel Application Program for Detecting Strabismus and Measuring Angle of Deviation.Clinical ophthalmology (Auckland, N.Z.) · 2026
- Inter-device Agreement of Device-Designated Chord Mu and Chord Alpha Measurements in Patients Undergoing Preoperative Cataract Evaluation.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- Diagnostic Utility of Accommodative Measurements by Nott Dynamic Retinoscopy in Children with Unilateral Amblyopia.Ophthalmic & physiological optics : the journal of the British College of Ophthalmic Opticians (Optometrists) · 2026
- Agreement between slit lamp retroillumination pictures and Scheimpflug retroillumination display for toric IOLs orientation.Journal of cataract and refractive surgery · 2026