International ophthalmology

Agreement between digital ocular alignment measurement devices and the prism alternate cover test in horizontal strabismus.

Orkun Özkan, Halil Hüseyin Çağatay, Bağdagül Koçak

Department of Ophthalmology, Faculty of Medicine, Uskudar University, Istanbul, Turkey. orkunozkan33@gmail.com. · Private Practice, Izmir, Turkey.

Observational · n=63

Key points

From the paper's conclusion
  • Digital ocular alignment devices demonstrated moderate agreement with PACT, particularly in patients with moderate-angle deviations. However, agreement decreased in larger deviations, and the wide limits of agreement indicate that individual measurements should be interpreted with caution. Therefore, PACT remains the clinical reference method for accurate deviation measurement, particularly in surgical planning.

Abstract

PURPOSE: To evaluate the agreement between digital ocular alignment devices and the prism alternate cover test (PACT) in patients with horizontal strabismus. METHODS: This retrospective cross-sectional study included 63 patients (31 esotropia, 32 exotropia). Ocular deviation was measured using the prism alternate cover test (PACT) at near (CTN) and distance (CTD) and compared with measurements obtained using the Welch Allyn Spot Vision Screener and the BCN Innova Sensor. Agreement was assessed using intraclass correlation coefficients and Bland-Altman analysis. Subgroup analyses were performed according to deviation magnitude (10-30 PD and >30 PD). RESULTS: The estimated ICC values indicated moderate agreement between the digital devices and PACT measurements. Agreement was generally stronger in exotropia than in esotropia. Bland-Altman analysis demonstrated small mean biases but wide limits of agreement. Agreement was more consistent within the 10-30 PD range and decreased in larger deviations. CONCLUSIONS: Digital ocular alignment devices demonstrated moderate agreement with PACT, particularly in patients with moderate-angle deviations. However, agreement decreased in larger deviations, and the wide limits of agreement indicate that individual measurements should be interpreted with caution. Therefore, PACT remains the clinical reference method for accurate deviation measurement, particularly in surgical planning.

Read the paper