Comparison of Cumulative Dissipated Energy Between Two Phacoemulsification Systems Using a Modified Flip-and-Chop Technique With Multidirectional Jerking Movement From a Phacoemulsification Handpiece.
Ming Chen, Hung-Yuan Lin, Emily J Kang
Ophthalmology, John A. Burns School of Medicine, University of Hawaii, Honolulu, USA. · Ophthalmology, Zhongli University Eye Center, Taoyuan, TWN.
Key points
From the paper's conclusion- The addition of this technique to the phacoemulsification handpiece may further improve the efficiency of phacoemulsification.
Abstract
Introduction The four-dimensional Unity phacoemulsification tip is designed to optimize tip motion and reduce time and energy consumption in cataract surgery. This study highlights a modified flip-and-chop technique with an added multidirectional jerking motion to the phacoemulsification handpiece intraocularly (referred to as the "five-dimensional flip-and-chop") and compares the cumulative dissipated energy (CDE) between the four-dimensional Unity system (Alcon Inc., Geneva, Switzerland) and the torsional Centurion system (Alcon Inc., Geneva, Switzerland) under this technique. Methods This is a retrospective, sequential, non-randomized comparison of 40 consecutive cases of Centurion to 40 consecutive cases of Unity performed by a single surgeon at a single ambulatory surgical suite. The initial cataract density between the groups was compared using optical coherence tomography (OCT) image quality (IQ) values from a single Maestro 2 device (Topcon, Tokyo, Japan). Results The two groups had comparable mean ages, sex distributions, and IQs. The mean CDE was significantly lower in the Unity group (5.39; SD=2.8) than in the Centurion group (8.23; SD=5.9), with p=0.0075. Conclusion Utilizing the modified "five-dimensional flip-and-chop" technique, the four-dimensional Unity system produced significantly lower CDE than the torsional Centurion system. The addition of this technique to the phacoemulsification handpiece may further improve the efficiency of phacoemulsification.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Clinical Comparison of 4D Phaco and Torsional Ultrasound Phacoemulsification Platforms: Association with Cumulative Dissipated Energy and Postoperative Inflammatory Response.Journal of clinical medicine · 2026
- Anterior Capsule Contrast and Capsulorhexis Geometry Across Visualization Systems in Cataract Surgery: An Exploratory Image-Based Comparison.Journal of clinical medicine · 2026
- Effect of phaco-tip orientation during phacoemulsification surgery (TIPS) on endothelial cell loss: Results of a randomized clinical trial.Acta ophthalmologica · 2026
- Effect of temperature variation on the corneal endothelial cell during femtosecond laser-assisted cataract surgery compared with conventional phacoemulsification cataract surgery: A prospective studyMedicine · 2026
- Efficiency and Safety of Torsional IP Mode and Combined Torsional-Longitudinal Mode for the Management of Grade 5 Cataract.Journal of the College of Physicians and Surgeons--Pakistan : JCPSP · 2026
- Low Versus High Intraoperative Intraocular Pressure During Phacoemulsification Cataract Surgery: A Systematic Review and Meta-Analysis of Randomized Controlled Trials.Journal of clinical medicine · 2026