Wearable, cuffless, and portable devices for blood pressure monitoring (2015-2025): a scoping review of technologies, calibration, clinical validation, and regulatory frameworks.
Juan Carlos Bustamante-Rodríguez, Jhosmer Ballena-Caicedo, Julio César Bautista-Zuta, Víctor Juan Vera-Ponce
Facultad de Medicina (FAMED), Universidad Nacional Toribio Rodríguez de Mendoza de Amazonas (UNTRM), Chachapoyas, Peru.
Key points
From the paper's conclusion- The 2015-2025 ecosystem shows rapid progress and a transition toward cuffless-specific validation. However, clinical utility depends on intended use, population, calibration/recalibration, and setting. Performance should not be extrapolated across technologies, populations, or contexts.
Abstract
BACKGROUND: Out-of-office blood pressure (BP) monitoring is essential for diagnosing and controlling hypertension. Between 2015 and 2025, wearable devices and portable cuffless platforms emerged that estimate BP without a conventional brachial cuff; their evaluation is complicated by technological heterogeneity, calibration, sensitivity to dynamic conditions, and recent device-specific standards. OBJECTIVE: To map the evidence published between 2015 and 2025 on wearable and portable cuffless devices for BP measurement or estimation, describing form factors, technologies, calibration/recalibration, validation, populations, settings, and standards or regulatory frameworks. METHODS: A scoping review was conducted in accordance with PRISMA-ScR and the Joanna Briggs Institute, using the PCC framework. Sources were searched in PubMed/MEDLINE, Scopus, Web of Science, and Embase, supplemented by device-directed searches and citation chasing. Extraction captured population, device, technology, setting, calibration, standard/protocol, comparator, metrics, limitations, conflicts, and funding. RESULTS: Sixty-three sources were included: 7 standards/consensus/regulatory documents, 1 case study, 6 reviews, and 49 primary studies. Primary evidence covered watches with miniaturized cuffs, wearable or portable cuffless devices, and rings/finger devices. Oscillometric wrist watches were methodologically closer to conventional automated monitors, although limited by positioning requirements, reading errors, and scarce ambulatory evidence. Cuffless devices relied on PPG/PWA, PTT/PAT, tonometry, or multimodal approaches, with calibration/recalibration and proprietary algorithms. Critical gaps persisted in post-calibration stability, dynamic conditions, sleep, treatment, arrhythmias, and special populations. CONCLUSIONS: The 2015-2025 ecosystem shows rapid progress and a transition toward cuffless-specific validation. However, clinical utility depends on intended use, population, calibration/recalibration, and setting. Performance should not be extrapolated across technologies, populations, or contexts.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Cuffless Devices for the Measurement of Blood Pressure: A Scientific Statement From the American Heart AssociationHypertension · 2025
- Wearable devices in the management of hypertension: Navigating the intersection between digital advancement and evidence-based practiceClinical Medicine · 2026
- An Unfinished but Necessary Path: Ring-Based Cuffless Blood Pressure Monitoring at the Crossroads of Innovation and Clinical Reality.Pulse (Basel, Switzerland) · 2026
- To be (IIb), or not to be (IIb) - That is the current status of cuffless blood pressure monitoring device.Clinical hypertension · 2026
- Evaluation of Blood Pressure and Heart Rate Measurements from Smartwatch HUAWEI WATCH D2 Data: A Preliminary Report.Healthcare (Basel, Switzerland) · 2026
- Validation of a wrist-worn dual-sensor applanation tonometry device for blood pressure measurement under rest and stress conditionsEuropean Heart Journal · 2026