Sen Bing, Mao-Hsiang Huang, Hung Cao, J.-C. Chiao
This work presents a noninvasive imaging method to locate veins using a tuned microwave loop resonator. It offers a low-cost, fast, and effective solution to the challenges in venipuncture. The sensor features a loop resonator with a 5.2 mm radius, incorporating a self-tuning mechanism, and operates at 2.408 GHz with a reflection coefficient of −48.77 dB. It generates localized high-intensity electric fields that penetrate tissues to sufficient depths, enabling the detection of veins based on shifts in resonant frequencies that are induced by the varied dielectric properties of blood vessels. Two-dimensional raster scan simulations of the cephalic and median cubital veins yielded a ∼25 MHz downward resonant-frequency shift between vein and non-vein positions, with the median cubital vein still detectable at depths up to 6 mm. To quantify generalization to real tissues, a decision tree classifier trained on 63 simulation samples and evaluated on 335 in vivo measurements achieved 82.09% classification accuracy (sensitivity 81.25%, specificity 83.02%), demonstrating that the simulation-derived frequency contrast transfers reliably to experimental data despite inter-subject tissue variability. Extensive tests conducted demonstrate the sensor’s effectiveness, producing consistent and distinguishable frequency shifts when the sensor moves on the skin across veins. This technology holds significant promise for improving venipuncture accuracy, minimizing complications, and enhancing patient comfort.