Within-Session Intra-Observer Reproducibility of Spleen Stiffness Measured by Endoscopic Ultrasound Point Shear-Wave Elastography: A Prospective Single-Operator Study.

Bogdan Miutescu, Adrian Burdan, Eyad Gadour, Felix Bende, Tudor Voicu Moga, Iulia Ratiu, Calin Burciu, Ana-Maria Ghiuchici,

Division of Gastroenterology and Hepatology, Department of Internal Medicine II, Victor Babes University of Medicine and Pharmacy, 300041 Timisoara, Romania. · Advanced Regional Research Center in Gastroenterology and Hepatology, Victor Babes University of Medicine and Pharmacy, 300041 Timisoara, Romania.

Observational · n=58

Key points

From the paper's conclusion
  • Therefore, the IQR/median <30% threshold requires independent validation before it can be recommended for clinical use.

Abstract

Background/Objectives: Endoscopic ultrasound point shear-wave elastography (EUS-pSWE) enables quantitative assessment of spleen stiffness, but the reproducibility of repeated measurements remains poorly established. This study evaluated within-session intra-observer reproducibility and the influence of acquisition-quality indicators. Methods: In this prospective, single-center study, 58 patients underwent ten consecutive splenic EUS-pSWE measurements performed by one operator without echoendoscope repositioning. Agreement between the medians of the first and last five measurements was assessed using intraclass correlation coefficients (ICC) and Bland-Altman analysis. Measurement quality was evaluated using velocity signal-to-noise ratio (VSN) and interquartile range-to-median ratio (IQR/median). Results: Overall, 66.6% of individual measurements achieved VSN > 60%, while only 20.7% of measurement series fulfilled the IQR/median < 30% criterion. Overall reproducibility was moderate (ICC 0.66, 95% CI 0.49-0.78), with a mean bias of -0.6 kPa and limits of agreement from -15.9 to 14.8 kPa. Among the 12 series meeting the ten-reading IQR/median < 30% criterion, ICC was 0.92 (95% CI 0.77-0.98). However, when both five-reading blocks separately met this criterion (n = 18), ICC was 0.61 (95% CI 0.23-0.83). Conclusions: Splenic EUS-pSWE demonstrated moderate overall within-session reproducibility. The higher agreement observed after whole-series quality-based selection was not consistently reproduced using an alternative block-level criterion. Therefore, the IQR/median < 30% threshold requires independent validation before it can be recommended for clinical use.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Within-Session Intra-Observer Reproducibility of Spleen Stiffness Measured by Endoscopic Ultrasound Point Shear-Wave Elastography: A Prospective Single-Operator Study — Science Skim