Quantitative Shear Wave Elastography: A Phantom—Based Comparison of Two Ultrasound Systems
Wadhhah Aldehani, Sarah Savaridas, Luigi Manfredi
University of Dundee
Key points
From the paper's conclusion- All phantom inclusions (8/8) measured below their assigned ground truth stiffness on both systems, with systematic underestimation ranging from 0.33 kPa to 109.57 kPa, indicating parameter-dependent measurement distortion linked to inclusion size, depth, and stiffness.
Abstract
To evaluate cross-platform measurement consistency and diagnostic threshold requirements in shear wave elastography (SWE), this study presents a robotically controlled, phantom-based validation framework to quantify and interpret inter-vendor variability that limits clinical standardisation. A custom-fabricated polyvinyl chloride-graphite phantom containing eight spherical inclusions (15–25 mm diameter, 25–95 mm depth, 23.53–259.58 kPa stiffness), representing breast tissue mechanical properties, was evaluated using Samsung HS50 and Aixplorer ultrasound systems. Robotic automation standardised probe positioning and contact, eliminating operator-dependent variability and enabling direct, system-level comparison. Cross-platform reproducibility, accuracy against mechanically validated ground truth, and diagnostic threshold performance were assessed across 80 measurements. Both systems demonstrated excellent intra-machine reproducibility (coefficient of variation: Samsung 0.42%, Aixplorer 0.46%) with strong inter-machine correlation (r = 0.9951, p < 0.0001). However, systematic bias of 7.05 kPa and 95% limits of agreement spanning 38.7 kPa revealed substantial cross-platform measurement differences. All phantom inclusions (8/8) measured below their assigned ground truth stiffness on both systems, with systematic underestimation ranging from 0.33 kPa to 109.57 kPa, indicating parameter-dependent measurement distortion linked to inclusion size, depth, and stiffness. Dynamic range compression was observed (Samsung: 68.7%, Aixplorer: 59.1% of true phantom range), providing a mechanistic explanation for diagnostic threshold instability. This study contributes an interpretable validation methodology that links SWE measurement bias to physical lesion properties and imaging system characteristics, rather than relying on correlation alone. Despite strong reproducibility, the observed system-dependent bias demonstrates that SWE measurements are not directly transferable across ultrasound platforms, and system-specific collaboration is required to ensure reliable clinical interpretation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Shear Wave Elastography for Characterization of Breast Lesions in Clinical RoutineJournal of Ultrasound in Medicine · 2026
- Quantitative Shear-Wave Elastography in Prostate Cancer: Lesion-Level Stiffness Thresholds and Association with Histopathological Aggressiveness.Cancers · 2026
- Within-Session Intra-Observer Reproducibility of Spleen Stiffness Measured by Endoscopic Ultrasound Point Shear-Wave Elastography: A Prospective Single-Operator Study.Diagnostics (Basel, Switzerland) · 2026
- Diagnostic performance of quantitative shear-wave elastography for reclassifying BI-RADS category 4 breast lesions: a retrospective single-center study.Frontiers in public health · 2026
- Effects of operator experience and measurement depth on skeletal muscle shear wave elastography.Journal of ultrasound · 2026
- Identification of local nonlinear shear-wave dynamics using Volterra kernels: application to prostate cancer detection.IEEE transactions on medical imaging · 2026