Effects of operator experience and measurement depth on skeletal muscle shear wave elastography.

Tetsuo Ota, Hina Yoshida, Mitsugu Yoneda

Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, Ishikawa, 920-0942, Japan. · Tono Chubu Medical Center, 1078-200 Asano, Hida Cho, Toki, Gifu, 509-5114, Japan.

Key points

From the paper's conclusion
  • These findings may contribute to the standardization of skeletal muscle SWE.

Abstract

PURPOSE: To investigate the effects of operator experience and measurement depth on shear wave elastography (SWE) measurements of skeletal muscle. Specifically, this study evaluated the reliability and agreement of SWE measurements obtained by experienced and inexperienced operators and examined whether patterns of shear wave velocity (SWV) across the region of interest (ROI) differed among muscles. METHODS: Twenty healthy young adults (40 limbs) participated in the study. SWE measurements were obtained from the biceps brachii (BB), triceps brachii (TB), rectus femoris (RF), and biceps femoris (BF) using a diagnostic ultrasound system. Two operators with different levels of ultrasound experience performed the measurements. Intra- and inter-rater reliability were assessed using intraclass correlation coefficients (ICCs) and Bland-Altman analysis. SWV was compared among superficial, intermediate, and deep positions within a fixed-size ROI; these positions were defined relative to the ROI rather than by absolute tissue depth from the skin surface. RESULTS: Intra-rater reliability was good to excellent (ICC (1,3) = 0.767-0.962), while inter-rater reliability ranged from moderate to excellent (ICC (2,2) = 0.706-0.956). Bland-Altman analysis revealed no fixed or proportional biases. Significant positional differences were found in the BB, TB, and RF, but not in the BF. SWV was higher at deeper ROI positions in the BB and TB, whereas the RF showed lower SWV in the deep position. CONCLUSIONS: Comparable reliability was observed between experienced and inexperienced operators. The pattern of SWV across the ROI differed among muscles, indicating muscle-specific regional variation in SWE measurements. These findings may contribute to the standardization of skeletal muscle SWE.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers