B-239 Assessment of a novel PBRTQC approach based on Kullback-Leibler Divergence
Claire Auger, Thando Gcingca, Merih Tesfazghi, Roa Harb
Key points
From the paper's conclusion- We present a novel approach to PBRTQC by using the Kullback-Leibler (KL) divergence.
Abstract
Background Patient based real-time QC (PBRTQC) can complement standard iQC methodologies with advantages including commutability and continuous versus interval-based monitoring. We present a novel approach to PBRTQC by using the Kullback-Leibler (KL) divergence. KL is essential to the cross-entropy loss function, widely used in machine learning model training to measure how closely predicted probability distribution matches the true data distribution. Here we harness a moving KL (MovKL) to identify shifts in data from simulated systematic error and sporadic low fliers. We test this against real sodium data and compare to other PBRTQC approaches. Methods 12542 sodium results measured on an Abbott Architect instrument were obtained. A standard block size of 25 was used. A “true” distribution was determined for MovKL from the first 100-500 samples. Bins were based on quintiles (<20%, 20-40%, 40-60%, 60-80%, >80%). Control limits set at the mean ?3.09 standard deviation (SD) of samples 100 to 500 were established for MovKL and the moving average (MA). For moving SD (MovSD), a standard approach was used for cut-off determination on the same “in-control” set. A “flag” was recorded when a second sample fell outside control limits within 6-100 samples of the first. A false-flagging rate was determined based on the frequency of flagging verified results without known error. A systematic bias (-10% to 10%) was applied to 2000 random samples across 100 iterations to estimate the median reported samples to error detection. Each method was tested on a real low-flier incident, determining number of samples to detection. Results Initial assessment of baseline “false-flagging” resulted in a total of .56% (70/12417), .88% (109/12417), and 11.95% (1484/12417) flags for MovKL, MA, and MovSD. When tested against systematic bias, stable CIs occurred at around 8% bias. Median samples required before detection occurred at 11 (7.95-16.05; MovSD), 12 (11-13; MA), and 20 (20-21; MovKL). This suggests MovKL may not identify bias shifts as quickly as MovAvg and MovSD. Each method was tested against a real-life sodium “low-flier” incident. MovSD, and MovKL identified error by the 27th, 32nd, and 36th overall sample and the 10th,14th, and 16th low-flier sample (decrease >2), respectively. While neither MovKL or MA flagged corrected sodium values, MovSD flagged the data erroneously. Conclusion Overall, these findings suggest that MA performed the best when considering both specificity and sensitivity. While MovKL performed slightly worse on sensitivity, its specificity was slightly improved over MA. MA methods are known to perform well on normally distributed analytes, like sodium. As MovKL relies on binning, it may not be as sensitive to normality assumptions and therefore be better suited for non-normally distributed analytes. Future work will focus on improving the sensitivity and evaluating the performance of MovKL in less normally distributed analytes.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Patient-Based Real-Time Quality Control for Semiquantitative Tests Based on Hierarchical Differences.Annals of laboratory medicine · 2026
- B-252 Prospective real-time implementation and clinical validation of patient-based real-time quality control using average of normals in a tertiary hospital laboratorClinical Chemistry · 2026
- A Two-Tier Patient-Based Real-Time Quality Control Program Integrating K-Means Department Stratification and Inter-Instrument Ratios Improves Detection of Small Systematic Errors in Creatinine Testing.Diagnostics (Basel, Switzerland) · 2026
- Integration of Patient-Based Real-Time Quality Control with Conventional Internal Quality Control: Improved Error Detection and Cost-Benefit Analysis in a Clinical Laboratory.EJIFCC · 2026
- Clustering of patient-derived laboratory data: A privacy-preserving distributed approach to enhance clinical laboratory quality control and regional external quality assessment.PloS one · 2026
- Statistical Monitoring of Multivariate Count Data From Real‐Time Particle CountersQuality and Reliability Engineering International · 2026