A Two-Tier Patient-Based Real-Time Quality Control Program Integrating K-Means Department Stratification and Inter-Instrument Ratios Improves Detection of Small Systematic Errors in Creatinine Testing.

Xuemei Wei, Ziqing Jiang, Jingzhu Nan, Zefeng Yan, Zhu Lin, Rong Li, Jie Feng, Yanhong Gao

Department of Clinical Laboratory, The First Medical Center of Chinese PLA General Hospital, Beijing 100853, China. · Department of Clinical Laboratory, Beijing Anzhen Hospital, Capital Medical University, Beijing 100029, China.

Key points

From the paper's conclusion
  • Conclusions: Integrating department stratification with ratio-based PBRTQC markedly improves detection of small systematic errors in creatinine testing, albeit with more false positives, offering a transferable framework for laboratories using multiple same-platform measuring systems.

Abstract

Background/Objectives: Patient-based real-time quality control (PBRTQC) complements traditional internal quality control. This study validates a ratio-based PBRTQC integrating department clustering with inter-instrument ratio monitoring. Methods: Creatinine data (384,730 cases, 2025) from three measuring systems (A, B, C) were analyzed; conventional PBRTQC on unstratified Instrument A data served as comparator. Departments were stratified by K-means clustering of annual mean/SD, keeping the low-mean cluster (124 departments). The data of each measuring system were split chronologically: first-layer PBRTQC was trained on the first part, the second trained the ratio programs, and the third provided independent validation. Three systematic errors (fixed additive offsets of +5.28 and +10.56 μmol/L, and a progressive offset rising to +10.56 μmol/L, corresponding to 6% and 12% of the first-layer action limit) were simulated to assess the true positive rate (TPR), false positive rate (FPR), median samples to detection (MNPed), detection time and diagnostic odds ratio (DOR). Results: Under the fixed +5.28 μmol/L (6%) offset, A-B TPR was 28.83% versus 6.77% for traditional PBRTQC, with MNPed reduced from 34 to 20. Under the progressive offset rising to +10.56 μmol/L (12%), A-B TPR was 29.00% versus 8.35%, and detection time fell from 20 to 14 min. Under the fixed +10.56 μmol/L (12%) offset, A-B and A-C TPR reached 70.15% and 74.72% (MNPed 8 and 5; detection time 2 and 1 min). Across all scenarios the ratio programs achieved higher DOR. A stratified single-instrument comparator showed complementary contributions of stratification and ratio monitoring (TPR at the +5.28 μmol/L offset: 6.77%, 15.61% and 28.83%). Conclusions: Integrating department stratification with ratio-based PBRTQC markedly improves detection of small systematic errors in creatinine testing, albeit with more false positives, offering a transferable framework for laboratories using multiple same-platform measuring systems.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers