Repeatability of Swan600 and Its Agreement with Lenstar LS900 for Ocular Biometry in Children.
Yuanbin Li, Yunyun Sun, Weibin Chen, Jing Fu
Beijing Ophthalmology & Visual Sciences Key Laboratory, Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China.
Key points
From the paper's conclusion- Conclusions: The Swan SW600 demonstrates high repeatability and close agreement with the Lenstar LS900 for AL, corneal radii, and AL/CR in children and adolescents, including preschool-aged children; however, the weaker agreement observed for LT and particularly WTW suggests that these parameters should be interpreted with caution.
Abstract
Background/Objectives: To evaluate the repeatability of the Swan SW600 optical biometer and its agreement with the Lenstar LS900 in measuring ocular biometric parameters in children and adolescents. Methods: This prospective study included 778 children and adolescents aged 3-18 years. Parameters included axial length (AL), corneal radii in the flattest and steepest meridians (R1 and R2), central corneal thickness (CCT), anterior chamber depth (AD), lens thickness (LT), white-to-white distance (WTW), and axial length-to-corneal radius ratio (AL/CR). Repeatability of Swan SW600 was assessed using within-subject standard deviation (Sw), coefficient of repeatability (CoR), coefficient of variation (CoV), and intraclass correlation coefficient (ICC). Agreement between devices was evaluated using paired t tests, ICC, and Bland-Altman analysis. Results: Swan SW600 repeatability was excellent for AL, R1, R2, CCT, AD, and AL/CR (ICC = 0.9918-0.9999) and lower for LT (ICC = 0.9060) and WTW (ICC = 0.8452). Agreement with Lenstar LS900 was strongest for AL (mean difference -0.050 ± 0.041 mm; 95% LoA -0.131 to 0.031 mm; ICC = 1.000), R1 (ICC = 0.986), R2 (ICC = 0.980), and AL/CR (ICC = 0.993). CCT and AD showed good agreement (ICC = 0.877 and 0.895), whereas LT showed lower agreement (ICC = 0.689). WTW showed poor agreement (ICC = 0.288). Age-stratified analyses showed consistently high agreement for AL, R1, R2, and AL/CR across all age groups, including children aged 3-6 years, whereas agreement for CCT, AD, LT, and WTW was more variable. Conclusions: The Swan SW600 demonstrates high repeatability and close agreement with the Lenstar LS900 for AL, corneal radii, and AL/CR in children and adolescents, including preschool-aged children; however, the weaker agreement observed for LT and particularly WTW suggests that these parameters should be interpreted with caution.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Clinical validation of a multifunction ocular biometer in children.Clinical & experimental optometry · 2026
- The Repeatability and Agreement of Two SS-OCT Optical Biometers in Adult Eyes Undergoing Cataract Surgery.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- Evaluation of axial length measurement and other biometric parameters using three different generations of biometers.Frontiers in medicine · 2026
- Agreement of Anterior Segment Biometrics Measured Using Image-Registered Swept-Source OCT and High-Frequency Ultrasound Biomicroscopy.Journal of glaucoma · 2026
- Extending a Standard SD-OCT Device to Ocular Biometry with Additional Software: Agreement and Clinical Applicability Compared with Ultrasound and Optical Biometry.Journal of clinical medicine · 2026
- Repeatability of an SD-OCT biometer (MS-39 AL) and agreement with a validated biometer (IOLMaster 700).Journal of cataract and refractive surgery · 2026