Leveraging Patient Data for Real-Time Inter-Instrument Comparison: Performance of the EWMA-Based PBRTQC Model on Triglyceride Measurements.
Xinyan Yao, Yuxin Wu, Hongyun Yang, Jie Guan
Department of Clinical Laboratory, Peking University First Hospital, Beijing, China. · Department of Laboratory Medicine, State Key Laboratory of Complex, Severe, and Rare Diseases, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences, Beijing, China.
Key points
From the paper's conclusion- PBRTQC using the EWMA algorithm enables continuous tracking of cross-instrument analytical performance. It serves as an effective adjunct for evaluating inter-instrument variations and strengthening multi-campus quality control.
Abstract
BACKGROUND: Patient-based real-time quality control (PBRTQC) offers cost-effective, continuous monitoring, but its application for multi-instrument harmonization within large healthcare systems remains underexplored. This study evaluated the feasibility of using PBRTQC for inter-instrument comparison using triglyceride (TG) measurements. METHODS: From 286,855 patient TG results across two campuses of Peking University First Hospital, 245,299 valid records were extracted. PBRTQC modeling utilized the exponentially weighted moving average (EWMA) algorithm. In parallel, fresh plasma samples from 60 patients covering a wide concentration range were collected, and inter-instrument comparison was carried out on six biochemical analyzers following CLSI EP9-A2 guidelines. The relative differences derived from EWMA means were compared with those from two-level conventional internal quality control (IQC) (Levels 1 and 2) on daily, weekly, and monthly bases, using half the total allowable error (TEa = ±7%) as the acceptability threshold. RESULTS: Big-data-driven comparison yielded relative differences for five comparator analyzers against the reference system ranging from -2.05% to 2.39%. The conventional EP9-A2 comparison using 60 fresh plasma samples also showed excellent correlations (r ≥ 0.9991), with all relative differences within the predefined acceptable limit. Over daily, weekly, and monthly trends, the EWMA mean trajectories closely paralleled those of conventional IQC means. These findings indicate that EWMA-derived conclusions align well with fresh specimen assessments, providing a robust method for monitoring cross-instrument analytical comparability. CONCLUSIONS: PBRTQC using the EWMA algorithm enables continuous tracking of cross-instrument analytical performance. It serves as an effective adjunct for evaluating inter-instrument variations and strengthening multi-campus quality control.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Integration of Patient-Based Real-Time Quality Control with Conventional Internal Quality Control: Improved Error Detection and Cost-Benefit Analysis in a Clinical Laboratory.EJIFCC · 2026
- Illustrative examples and data from patient-based real-time quality controlThe Clinical Biochemist Reviews · 2026
- Exploring patient-based real-time quality control: Insights from EWMA and simulation modeling.Clinica chimica acta; international journal of clinical chemistry · 2026
- B-252 Prospective real-time implementation and clinical validation of patient-based real-time quality control using average of normals in a tertiary hospital laboratorClinical Chemistry · 2026
- Clustering of patient-derived laboratory data: A privacy-preserving distributed approach to enhance clinical laboratory quality control and regional external quality assessment.PloS one · 2026
- Patient-Based Real-Time Quality Control for Semiquantitative Tests Based on Hierarchical Differences.Annals of laboratory medicine · 2026