Analytical Agreement Between the Vision C Automated ESR Analyzer and the Manual Westergren Reference Method: A Prospective Method Comparison Study.
Ayça Tuzcu, Kenan Yörük, Yusuf Kurtulmuş, Mustafa Yılmaz
Department of Biochemistry, Faculty of Medicine, Aydin Adnan Menderes University, 09100 Aydin, Türkiye.
Key points
From the paper's conclusion- Conclusions: The Vision C analyzer showed a significant, level-dependent negative bias versus Westergren, widening at higher ESRs; strong correlation and RCPA TEa compliance across the full range, including the high subgroup, support its suitability for routine use.
Abstract
Background/Objectives: The erythrocyte sedimentation rate (ESR) is a widely used biomarker of systemic inflammation, with established roles in rheumatologic, infectious, and malignant disease. The Westergren method remains the reference procedure, but automated analyzers are increasingly used clinically, and methodological differences may introduce systematic bias. This study evaluated analytical agreement between the Vision C automated ESR analyzer (EDTA) and the Westergren reference (sodium citrate). Methods: This prospective, single-center study followed CLSI EP09-ED3. A total of 228 consecutive adult patients were enrolled. Agreement was assessed by Bland-Altman analysis, Passing-Bablok regression, and Pearson correlation, with subgroup analyses (<40, 40-80, >80 mm/h) and against RCPA allowable total error (TEa) specifications. Results: Bland-Altman analysis showed a significant mean bias of -3.30 mm/h (95% CI: -4.25 to -2.35), with increasing dispersion at higher ESR values indicating heteroscedasticity; regression-based limits of agreement therefore widened progressively with ESR magnitude. Passing-Bablok regression showed no proportional bias but a significant constant bias; subgroup regression was not performed due to truncated ranges. Correlation was strong overall (r = 0.961, p < 0.001) and declined across subgroups (r = 0.821, 0.740, 0.687), consistent with range restriction rather than reduced agreement. The analyzer met RCPA TEa criteria across both ranges (≤20 mm/h: bias -2.55 mm/h within ±6.0 mm/h; >20 mm/h: percentage bias -10.49% within ±30%), including the high-ESR subgroup (approximately 6-9%). Conclusions: The Vision C analyzer showed a significant, level-dependent negative bias versus Westergren, widening at higher ESRs; strong correlation and RCPA TEa compliance across the full range, including the high subgroup, support its suitability for routine use. Wide regression-based limits and the limited high subgroup sample size (n = 20) warrant caution and further study but do not support routine manual confirmation, since predefined acceptance criteria are met. Method-specific reporting and consistent instrument use in follow-up are recommended.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of Westergren and Automated Erythrocyte Sedimentation Rate Measurements in Dogs and Cats.Animals : an open access journal from MDPI · 2026
- Discordance between the erythrocyte sedimentation rate and C-reactive protein in rheumatic diseases: reported frequencies, patterns, and clinical interpretation.Frontiers in medicine · 2026
- A exploratory pilot method comparison study of two viscoelastic coagulation analyzers.Frontiers in medicine · 2026
- Method Comparison of PT, INR, and aPTT Results Obtained Using Two Coagulation Analyzers in Routine Laboratory Practice.Diagnostics (Basel, Switzerland) · 2026
- Measuring Erythrocyte Sedimentation Rate Using Photometric Rheology Technology Demonstrates Superior Sample Stability as Compared to the Westergren Method.Diagnostics (Basel, Switzerland) · 2026
- Numerical and Interpretive Comparability of Optical and Mechanical Coagulation Analyzers: A Zlog-Supported Method Comparison Study.Diagnostics (Basel, Switzerland) · 2026