Numerical and Interpretive Comparability of Optical and Mechanical Coagulation Analyzers: A Zlog-Supported Method Comparison Study.
Xinjian Cai, Wei Yang, Qiuxia Lu, Yiteng Lin
Department of Clinical Laboratory Medicine, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital & Shenzhen Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Shenzhen 518116, China.
Key points
From the paper's conclusion- Conclusions: Numerical and interpretive comparability are distinct dimensions.
Abstract
Background/Objectives: In laboratories using coagulation analyzers with different clot-detection principles and platform-specific reference intervals (RIs), numerical agreement does not necessarily ensure concordant clinical interpretation. We evaluated how analytical agreement and RI-based interpretation diverge between mechanical and optical coagulation analyzers. Methods: Paired duplicate measurements were performed on the STA R Max (Stago) and ACL TOP 750 LAS (Werfen) for prothrombin time (PT), activated partial thromboplastin time (APTT), thrombin time (TT), fibrinogen (FIB), international normalized ratio (INR), and antithrombin (AT). Numerical agreement was assessed by Passing-Bablok regression and log-ratio Bland-Altman analysis against predefined total allowable error (TEa) limits. Each result was standardized using its platform-specific RI, and the between-platform zlog difference (Δzlog) described differences in RI-relative position. Results: Stago yielded higher values than Werfen for PT, APTT, TT, and FIB, with mean biases of +15.6% to +20.4%, whereas INR and AT showed close agreement (-1.1% and +1.6%). Interpretive concordance among the primary assays ranged from 61.9% for APTT to 89.2% for FIB. The analytical-by-interpretive matrix identified specimens that were within TEa but classified differently and specimens that were beyond TEa but retained the same RI classification. Median Δzlog values were -0.43 for PT, +0.09 for APTT, -0.66 for TT, and +0.77 for FIB. Conclusions: Numerical and interpretive comparability are distinct dimensions. Combining the analytical-by-interpretive matrix with Δzlog can reveal clinically relevant divergence not apparent from bias-versus-TEa assessment alone. In this purposively selected dataset, divergence arose mainly from RI offset for TT and analytical dispersion for APTT; the sample-specific discordance rates are not population prevalence estimates.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Method Comparison of PT, INR, and aPTT Results Obtained Using Two Coagulation Analyzers in Routine Laboratory Practice.Diagnostics (Basel, Switzerland) · 2026
- Method Comparison of Three Automated D-Dimer Assays.Diagnostics (Basel, Switzerland) · 2026
- Impact of reference interval source on result classification in pediatric coagulation testing.Clinica chimica acta; international journal of clinical chemistry · 2026
- A exploratory pilot method comparison study of two viscoelastic coagulation analyzers.Frontiers in medicine · 2026
- Instrument-specific reference intervals for coagulation factor XII and antithrombin III: A regional study in Jiangsu, China.Journal of immunological methods · 2026
- Analytical Agreement Between the Vision C Automated ESR Analyzer and the Manual Westergren Reference Method: A Prospective Method Comparison Study.Diagnostics (Basel, Switzerland) · 2026