Comparison of Westergren and Automated Erythrocyte Sedimentation Rate Measurements in Dogs and Cats.
Marisa Masucci, Alessandra Caprì, Alessandra Cartocci, Daniela Diamanti, Giulia Donato, Nicola Maria Iannelli, Carolina Pieroni, Flavia Rosace,
Department of Veterinary Science, University of Messina, 98168 Messina, ME, Italy. · ASC "I Periodeuti", 89122 Reggio Calabria, RC, Italy.
Key points
From the paper's conclusion- Healthy cats had significantly higher values than healthy dogs using both the Westergren and MINI-PET methods.
Abstract
The erythrocyte sedimentation rate (ESR) is a routine marker of inflammation in human medicine, and it is influenced by erythrocyte features and the level of some plasma proteins. The Westergren method is the reference technique in humans; however, this manual technique has been scarcely used in animals and has not been validated. An automated veterinary in-clinic device (MINI-PET) is able to measure ESR on EDTA-blood tubes. This study aimed to evaluate the comparability of ESR values measured with two tests in cats and dogs admitted to a veterinary clinic and in anemic individuals. Analytical comparability and decision limits were calculated in both species and clinical groups. Comparisons were also made between the two species. The two ESR methods had different decision limits in the investigated population of dogs and cats, and a method-specific interpretation of their results is suggested. Healthy cats had significantly higher values than healthy dogs using both the Westergren and MINI-PET methods. By both methods, anemic individuals had higher ESR values than the matched subgroups with erythrogram values within the reference range, probably due to concurrent inflammation in most anemic animals studied. In general, ESR measures should be evaluated in conjunction with erythrogram values.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Analytical Agreement Between the Vision C Automated ESR Analyzer and the Manual Westergren Reference Method: A Prospective Method Comparison Study.Diagnostics (Basel, Switzerland) · 2026
- Analytical validation of the Mindray BC-75R Vet hematology analyzer and comparison with the Siemens Advia 2120i for application in dogs and cats.Journal of veterinary diagnostic investigation : official publication of the American Association of Veterinary Laboratory Diagnosticians, Inc · 2026
- Comparison of Von Frey Filaments and Electronic Von Frey Device in Healthy dogs - a Prospective, randomized, experimental study.Veterinary journal (London, England : 1997) · 2026
- Reference intervals for myocardial T1, T2, and extracellular volume in clinically normal dogs and cats at 3-Tesla magnetic resonance imaging: a single-center studyAmerican Journal of Veterinary Research · 2026
- Blood coagulation of different animal species assessed with rotational thromboelastometry-Are there differences related to species, sex or age?Animal models and experimental medicine · 2026
- A-228 Multivariable Statistical Analysis Techniques to Identify Subpopulations in Method Comparisons Studies. A Practical Example Comparing with Two Methods for Erythrocytes Sedimentation Rate.Clinical Chemistry · 2026