A exploratory pilot method comparison study of two viscoelastic coagulation analyzers.
Daniel Ricardo Argumedo, Júlio Gallinaro Maranho, Martha R Ortiz-Posadas
Center of Natural and Human Science, Federal University of ABC, Santo André, Brazil. · Engineering, Modelling and Applied Social Sciences Center, Federal University of ABC, São Bernardo do Campo, Brazil.
Key points
From the paper's conclusion- This preliminary exploratory study evaluated the analytical performance of two different brand of viscoelastic coagulation analyzers in real-time blood samples from ten anonymous patients were analyzed on both devices, and the results were compared using Spearman correlation, Wilcoxon paired testing, Passing-Bablok regression, Bland-Altman agreement analysis, and Lin's concordance correlation coef
Abstract
This preliminary exploratory study evaluated the analytical performance of two different brand of viscoelastic coagulation analyzers in real-time blood samples from ten anonymous patients were analyzed on both devices, and the results were compared using Spearman correlation, Wilcoxon paired testing, Passing-Bablok regression, Bland-Altman agreement analysis, and Lin's concordance correlation coefficient. Systematic biases and variable concordance were observed across several parameters, indicating that the current exploratory dataset does not support direct numerical interchangeability between the two platforms without further testing in larger groups. As new platforms enter the global market, method-comparison studies using agreement-based statistical frameworks should be considered.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Method Comparison of PT, INR, and aPTT Results Obtained Using Two Coagulation Analyzers in Routine Laboratory Practice.Diagnostics (Basel, Switzerland) · 2026
- Numerical and Interpretive Comparability of Optical and Mechanical Coagulation Analyzers: A Zlog-Supported Method Comparison Study.Diagnostics (Basel, Switzerland) · 2026
- Method Comparison of Three Automated D-Dimer Assays.Diagnostics (Basel, Switzerland) · 2026
- Analytical Agreement Between the Vision C Automated ESR Analyzer and the Manual Westergren Reference Method: A Prospective Method Comparison Study.Diagnostics (Basel, Switzerland) · 2026
- A-249 Evaluation of three cartridge-based analyzers for viscoelastic testingClinical Chemistry · 2026
- A-231 Tri-Modal Detection in Coagulation Testing: Performance Evaluation of the Zybio EXT 9800 Coagulation Analyzer and Its Utility in Challenging Clinical SamplesClinical Chemistry · 2026