Magnetic Relaxation Measurements of Ex Vivo Human Liver Specimens Using a SERF Magnetometer.
Kexun Tang, Junhao Zheng, Yuxin Zhou, Dongzi Zhang, Chao Ren, Meijie Chen, Feidan Yu, Yi Ruan,
Key Laboratory of Quantum Precision Measurement of Zhejiang Province, School of Physics and Optical Engineering, Zhejiang University of Technology, Hangzhou 310023, China. · Key Laboratory of Multi-Omics Precision Diagnosis and Treatment of Liver Diseases, Department of General Surgery, Sir Run Run Shaw Hospital, School of Medicine, Zhejiang University, Hangzhou 310016, China.
Key points
From the paper's conclusion- We used a spin exchange relaxation free (SERF) atomic magnetometer to record magnetic relaxation signals from resected human liver tissue.
Abstract
Reliable assessment of liver iron concentration (LIC) is important for diagnosing and monitoring iron overload disorders. We used a spin exchange relaxation free (SERF) atomic magnetometer to record magnetic relaxation signals from resected human liver tissue. A total of 23 specimens from 20 patients, including normal, fatty, cirrhotic, and hepatocellular carcinoma (HCC) liver, were polarized with a Helmholtz coil inside a magnetic shield, and the field decay after switch off was recorded. For each specimen, a water-referenced, mass-normalized relaxation amplitude ΔB (pT/g) was derived. ΔB was compared with the HISTO-derived water transverse relaxation rates, R2water, from the patients' clinical MRI. For HCC, SERF measurements were made on resected tumor tissue, whereas MRI sampled non-cancerous liver parenchyma in the same patient. Within each category, ΔB increased with R2water and rank order agreed fully in three of the four categories. Slope differences between categories were significant with conventional standard errors but not with heteroscedasticity-consistent (HC3) standard errors and are therefore regarded as exploratory. These measurements establish a SERF workflow for comparing post-polarization magnetic relaxation amplitudes across resected human liver specimens.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Calibration of 3‐T MR Relaxometry for Quantitative Liver Iron Assessment Using Human Liver ExplantsNMR in Biomedicine · 2026
- Water-T1, PDFF, and T2* mapping of the liver using chemical shift-encoded MOLLI.Abdominal radiology (New York) · 2026
- Multiparametric 0.4-T low-field MRI mapping for metabolic dysfunction-associated steatotic liver disease in rats.European radiology experimental · 2026
- Mathematical Modeling of Liver Metabolic Activity Under Ex Vivo Conditions upon Exposure to Magnetic Nanoparticles.Biomedicines · 2026
- Inter-/Intra-Reader Variability in Ultrasound Measurement of Skin-to-Liver Capsule Distance and Hepato-Renal Index.Ultrasound in medicine & biology · 2026
- Segment-resolved assessment of hydration effects on liver water-specific T1 in comparison with T1-MOLLI and transient elastography.Quantitative imaging in medicine and surgery · 2026