Inter-/Intra-Reader Variability in Ultrasound Measurement of Skin-to-Liver Capsule Distance and Hepato-Renal Index.
Samay Prakash, Sai Dhanush Reddy Jeggari, Xiaohong Wang, Peng Guo, Michelle Luster, Emad Sorial, Saira Khaderi, Eugene Cheah,
Center for Ultrasound Research & Translation, Massachusetts General Hospital, Boston, MA, USA; Harvard Medical School, Boston, MA, USA. · Center for Ultrasound Research & Translation, Massachusetts General Hospital, Boston, MA, USA; Harvard Medical School, Boston, MA, USA. Electronic address: eaglep91@outlook.com.
Key points
From the paper's conclusion- SCD was repeatable and reproducible, supporting its viability as a quantitative ultrasound biomarker in the assessment of hepatic steatosis. HRI was imprecise, limiting its utility as a standalone metric. Measurement precision did not significantly differ among etiologies of cirrhosis.
Abstract
BACKGROUND: Metabolic dysfunction-associated steatotic liver disease (MASLD), if untreated, results in cirrhosis. Skin-to-liver capsule distance (SCD) and hepato-renal index (HRI) are ultrasound-derived biomarkers used in the quantitative assessment of hepatic steatosis, but manual measurements introduce operator variability. This study quantified inter- and intra-reader variability in SCD and HRI among patients with cirrhosis and assessed whether measurement precision differed by disease etiology. METHODS: A total of 1825 ultrasound examinations from 560 adults with cirrhosis were retrospectively identified. SCD and HRI reproducibility coefficients (RDC) were quantified between two independent readers; repeatability coefficients (RC) were quantified for repeated measurements by a single reader with a ≥28-d washout period. Precision was compared among MASLD, alcohol-related liver disease, metabolic dysfunction and alcohol-associated liver disease and chronic hepatitis C virus subgroups using one-way ANOVA with pairwise t-tests and false discovery rate (FDR) correction (p ≤ 0.05). RESULTS: A total of 1616 images were evaluable for SCD and 1065 for HRI. Median SCD was 21.2 mm and 20.8 mm; median HRI was 1.11 and 1.14, for Reader 1 and Reader 2, respectively. RDCs were 19.9% (SCD) and 55.8% (HRI) and RCs were 19.7% (SCD) and 52.2% (HRI). Neither SCD nor HRI reproducibility differed by etiology. HRI showed overall inter-reader variability differences (p = 0.023), but no pairwise comparison remained significant after FDR correction. CONCLUSION: SCD was repeatable and reproducible, supporting its viability as a quantitative ultrasound biomarker in the assessment of hepatic steatosis. HRI was imprecise, limiting its utility as a standalone metric. Measurement precision did not significantly differ among etiologies of cirrhosis.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- US Markers in the Noninvasive Assessment of Hepatic Steatosis in Metabolic Dysfunction-associated Steatotic Liver Disease.Radiology · 2026
- Advances in quantitative ultrasound for metabolic dysfunction-associated steatotic liver disease diagnosisFrontiers in Physiology · 2026
- Diagnostic performance of quantitative ultrasound techniques for the diagnosis and grading of hepatic steatosis in metabolic dysfunction-associated steatotic liver disease: a systematic review and network meta-analysis.BMC gastroenterology · 2026
- Endoscopic ultrasound strain elastography in MASLD: Evaluation against transient elastography and non-invasive markers.Hepatology forum · 2026
- Ultrasound Shear Wave Elastography for Liver Stiffness and Attenuation Imaging for Liver Steatosis in Patients With Metabolic Dysfunction-Associated Fatty Liver Disease.Medical science monitor : international medical journal of experimental and clinical research · 2026
- Ultrasound quantification of liver fat in MASLD.European radiology · 2026