Calibration of 3‐T MR Relaxometry for Quantitative Liver Iron Assessment Using Human Liver Explants
Petr Kordac, Milan Hajek, Petr Sedivy, Dita Pajuelo, Martin Kveton, Ondrej Fabian, Iva Subhanová, Ferdinand Los,
Key points
From the paper's conclusion- These findings support the feasibility of deriving local MR relaxometry calibrations from human liver explants, while further prospective validation against an independent LIC reference is required before clinical application.
Abstract
MR relaxometry has become the preferred noninvasive method for estimating hepatic iron from R2 and R2* relaxation rates. However, the precise quantification of liver iron concentration (LIC) requires calibration, and published equations relating relaxation rates to LIC have limited transferability across scanners and acquisition protocols. This study aimed to establish local 3‐T calibration equations relating R2 and R2* relaxation rates to chemically determined LIC using human liver explants and to evaluate their applicability in independent in vivo cohorts. Thirty‐six human liver explants underwent MR examination on a 3‐T system using multi‐echo spin‐echo and multi‐echo gradient‐echo sequences to obtain R2 and R2* relaxation rates. LIC was determined chemically by atomic absorption spectrophotometry. Calibration equations were derived using linear regression and compared with previously published models. The R2*‐based calibration was subsequently applied to independent in vivo cohorts examined using the same MR system and comparable acquisition protocols. Both MR methods correlated strongly with chemically determined LIC (R2*: r = 0.95, p < 0.001; R2: r = 0.92, p < 0.001). Compared with published models, the local calibrations demonstrated lower prediction errors and narrower limits of agreement within the studied explant cohort. Application of the R2* equation to in vivo datasets yielded physiologically plausible LIC estimates, with the lowest values observed in patients with primary sclerosing cholangitis, intermediate values in healthy controls and the highest values in subjects with elevated serum ferritin concentrations. Explant‐derived 3‐T calibration equations demonstrated strong association with chemically measured LIC and generated physiologically plausible LIC estimates in vivo. These findings support the feasibility of deriving local MR relaxometry calibrations from human liver explants, while further prospective validation against an independent LIC reference is required before clinical application.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Magnetic Relaxation Measurements of Ex Vivo Human Liver Specimens Using a SERF Magnetometer.Biosensors · 2026
- Clinical validation of T1ρ mapping for the assessment of hepatic fibrosis in patients with chronic liver diseaseEuropean Radiology · 2025
- Free-Breathing PDFF and R2* Quantification at 0.55T: Reduced Respiratory Motion Sensitivity and a Time-Resolved 4D MRI ApproacharXiv · 2026
- Multiparametric 0.4-T low-field MRI mapping for metabolic dysfunction-associated steatotic liver disease in rats.European radiology experimental · 2026
- Standardizing Liver MR Elastography: Guidelines for Reliable Stiffness Measurement in Chronic Liver Disease.Radiology · 2026
- Liver functional radiomics of gadoxetic acid-enhanced magnetic resonance imaging: A proof-of-concept studymedRxiv · 2026