Water-T1, PDFF, and T2* mapping of the liver using chemical shift-encoded MOLLI.
Amol Pednekar, Joshua S Greer, Pradipta Debnath, Andrew T Trout, Jonathan R Dillman, Mary Kate Manhard
Radiology, Cincinnati Children's Hospital Medical Center, Cincinnati, USA. Amol.Pednekar@cchmc.org. · MR Clinical Science, Philips (United States), Cincinnati, USA.
Key points
From the paper's conclusion- CSE-MOLLI enables simultaneous water-T1, PDFF, and T2* mapping of the liver within a single breath-hold. Phantom validation, repeatability assessment, and preliminary in vivo results in healthy, steatotic, and non-steatotic liver disease participants suggest that hepatic water-T1 measurements are less sensitive to fat-related deviations than conventional MOLLI-based T1 estimates. Further studies are needed to determine clinical utility in diffuse liver disease.
Abstract
PURPOSE: Hepatic MRI T1 relaxation time is a putative measure of fibrosis and inflammation but may be confounded by hepatic fat. This study evaluated a chemical shift-encoded Modified Look-Locker Inversion Recovery (CSE-MOLLI) technique for simultaneous water-specific T1 (water-T1), proton density fat fraction (PDFF), and T2* mapping within a single breath-hold. METHODS: Numerical simulations, reference standard phantoms, and in vivo imaging were used to assess CSE-MOLLI performance. Twenty participants, including healthy/asymptomatic controls, individuals with steatotic liver disease, and individuals with non-steatotic chronic liver disease, underwent CSE-MOLLI and comparator quantitative MRI sequences. All quantitative maps were generated in real-time on the scanner. Water-T1, PDFF, and T2* were measured from regions of interest (ROIs) by two readers and compared with reference standard methods using simple and intraclass correlation and Bland-Altman analysis. Scan-rescan repeatability was assessed for CSE-MOLLI. RESULTS: Phantom experiments demonstrated strong agreement between CSE-MOLLI and reference measurements for water-T1 (r = 0.98), PDFF (r = 1.00), and T2* (r = 0.99) estimates. In vivo, inter-reader agreement was excellent for water-T1, PDFF, and T2* (ICC > 0.97). CSE-MOLLI showed very strong correlation and excellent agreement with comparator PDFF (r = 1.00; ICC = 0.98) and water-T1 (r = 0.95; ICC = 0.89), while T2* demonstrated strong correlation and moderate agreement with comparator values (r = 0.67; ICC = 0.65). Water-T1 remained relatively consistent across PDFF values from 1% to 43%, suggesting reduced sensitivity of CSE-MOLLI to fat-related deviation. Water-T1 exhibited good scan-rescan repeatability (bias = 2.0 ms, ICC = 0.95). CONCLUSION: CSE-MOLLI enables simultaneous water-T1, PDFF, and T2* mapping of the liver within a single breath-hold. Phantom validation, repeatability assessment, and preliminary in vivo results in healthy, steatotic, and non-steatotic liver disease participants suggest that hepatic water-T1 measurements are less sensitive to fat-related deviations than conventional MOLLI-based T1 estimates. Further studies are needed to determine clinical utility in diffuse liver disease.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Cross‐Vendor Validation of Proton Density Fat Fraction and T₁ Mapping Using a Combined Proton Density Fat Fraction— T1 PhantomJournal of Magnetic Resonance Imaging · 2026
- Clinical utility of MOLLI T1 mapping using averaged in- and out-of-phase echo times (AIOP) for quantitative assessment of hepatic fibrosis in steatotic livers.The British journal of radiology · 2026
- Liver T1, T2, T2*, and PDFF Mapping at 0.55 T Using Rosette MRF With Optimized Sequence Design and Deep Image Reconstruction.Magnetic resonance in medicine · 2026
- Clinical validation of T1ρ mapping for the assessment of hepatic fibrosis in patients with chronic liver diseaseEuropean Radiology · 2025
- Segment-resolved assessment of hydration effects on liver water-specific T1 in comparison with T1-MOLLI and transient elastography.Quantitative imaging in medicine and surgery · 2026
- Superiority of magnetic resonance elastography and proton density fat fraction: an umbrella meta-analysis of non-invasive imaging for hepatic steatosis and fibrosis in metabolic dysfunction-associated steatotic liver disease and chronic hepatitis B.Quantitative imaging in medicine and surgery · 2026