Long-Duration Microdosimetry at the LNGS External Laboratory
Anna Bianchi, Stefania Canella, Alberto Fazzi, Anna Selva, Valeria Conte
Key points
From the paper's conclusion- This study assessed the feasibility of environmental microdosimetry in the external laboratory of the Laboratori Nazionali del Gran Sasso (LNGS) using a refurbished 5 cm spherical tissue-equivalent proportional counter (EuTEPC).
Abstract
Environmental radiation fields are characterized by very low dose rates and sparse energy-deposition events, conditions under which absorbed dose alone provides an incomplete description of irradiation at cellular and subcellular scales. This study assessed the feasibility of environmental microdosimetry in the external laboratory of the Laboratori Nazionali del Gran Sasso (LNGS) using a refurbished 5 cm spherical tissue-equivalent proportional counter (EuTEPC). The detector was converted into a stand-alone system, equipped for remote control and timed spectral acquisition, and installed at LNGS for long-duration measurements. Initial LNGS measurements were dominated by electromagnetic interference and did not yield physically interpretable spectra. Practical troubleshooting indicated that the dominant disturbance was associated with sporadic electromagnetic pickup, although the coupling mechanism was not investigated systematically. Extending the shielding to the detector and the complete cabling chain substantially reduced the noise and enabled the acquisition of stable lineal-energy distributions. Spectra acquired several months apart preserved the same overall shape, indicating consistent measurement performance under very low event-rate conditions. Comparison with measurements at Legnaro National Laboratories (LNLs) and with published environmental microdosimetric data from Ottawa and the high-altitude site of Gulmarg showed physically consistent spectral features, while also highlighting differences consistent with variations in radiation-field composition, detector characteristics and acquisition conditions. These results establish an operational platform for environmental microdosimetry at LNGS and identify electromagnetic compatibility as a critical requirement for measurements in very-low-background environments.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Integrated spatiotemporal assessment of ambient radiation fields in a nuclear medicine department using OSL and time-resolved Geiger-Müller monitoring.Journal of radiological protection : official journal of the Society for Radiological Protection · 2026
- Environmental radiological baseline characterization of a proposed medical cyclotron facility in Indonesia using dual-detector measurements and geostatistical analysis.Environmental geochemistry and health · 2026
- Site-specific field comparison of in situ NaI(Tl) gamma-ray spectrometry for ambient dose equivalent rate estimation and 137Cs spectral screening.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Urban Dose Rate Estimation Using Distributed Static D3S Gamma-Ray Spectrometers.Sensors (Basel, Switzerland) · 2026
- Characterizing long-term trends and spatiotemporal dynamics of ambient gamma radiation dose rates alongside exploratory meteorological factors.Journal of radiological protection : official journal of the Society for Radiological Protection · 2026
- National intercomparison of environmental gamma radiation dose rate in China based on ice surface cosmic ray response measurements.Journal of environmental radioactivity · 2026