Integrated spatiotemporal assessment of ambient radiation fields in a nuclear medicine department using OSL and time-resolved Geiger-Müller monitoring.
Kanokon Poonak, Poonnaphob Sopapan, Phiyada Thongkuam, Marisa Sureeyaphan, Ladawan Tanakitchaiyasit, Nopteera Sungsriphet, Panatsada Awikunprasert, Piyarat Parklug,
Department of Radiological Technology, Navamindradhiraj University, 681 Samsen Road, Wachira Phaya Subdistrict, Dusit District, Bangkok, 13000, Thailand. · Department of Radiology, Navamindradhiraj University, 681 Samsen Road, Wachira Phaya Subdistrict, Dusit District, Bangkok, 13000, Thailand.
Key points
From the paper's conclusion- This study integrated area-mounted optically stimulated luminescence (OSL) dosimetry with time-resolved Geiger-Müller (GM) survey monitoring to characterize both spatial dose distribution and temporal dose-rate behavior in a clinical nuclear medicine department.
Abstract
Occupational radiation protection in nuclear medicine is challenged by ambient radiation fields that vary both spatially and temporally. This variability arises from routine radiopharmaceutical handling and injected patients, who become temporary mobile radiation sources. Conventional surveillance based on passive dosimetry provides cumulative information but cannot resolve short-term operational fluctuations. This study integrated area-mounted optically stimulated luminescence (OSL) dosimetry with time-resolved Geiger-Müller (GM) survey monitoring to characterize both spatial dose distribution and temporal dose-rate behavior in a clinical nuclear medicine department. Monthly ambient dose equivalent, H*(10), was measured at 34 positions grouped into 14 functional areas across controlled, supervised, and uncontrolled zones, while ambient dose rate was recorded at 1-s intervals during routine working hours (08:00-16:00 h) over four weeks. Estimated annual effective dose (AED) and excess lifetime cancer risk (ELCR) were derived from the OSL data. The syringe placement area showed the highest monthly H*(10) (0.405 mSv month⁻¹), followed by the patient waiting room (0.288 ± 0.214 mSv month⁻¹) and injection room (0.204 ± 0.101 mSv month⁻¹). The patient waiting room also recorded the highest four-week mean dose rate (1.21 ± 0.24 µSv h⁻¹). Functional-area AED indicators ranged from 1.18 to 5.30 mSv y⁻¹, indicating location-specific variation in the annualized exposure estimates. The between-zone difference in H*(10) was not significant (p = 0.262), whereas variation among controlled-area locations was significant (p = 0.046). Time-resolved monitoring further revealed transient dose-rate peaks not captured by cumulative dosimetry. Integrated monitoring may provide a more complete basis for spatial prioritization and workflow-informed optimization in nuclear medicine workplaces.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Environmental radiological baseline characterization of a proposed medical cyclotron facility in Indonesia using dual-detector measurements and geostatistical analysis.Environmental geochemistry and health · 2026
- Comparing monitored occupational exposure and unmonitored indoor radon in healthcare settings.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Long-Duration Microdosimetry at the LNGS External LaboratoryApplied Sciences · 2026