Evaluating Analytical Performance of Commercial Screen-Printed Electrodes for Point-of-Care Uric Acid Sensing: A Comparative Study
Ligia Chelmea, Mihaela Badea, Sehrish Bilal, Davide Calin Mereu, Andrea Elena Neculau, Radu Dan Necula, Maria Elena Cocuz, Ovidiu Gabriel Bratu,
Key points
From the paper's conclusion- Here, we systematically compare two commercially available SPEs, DRP-110 and DRP-150, for UA detection using differential pulse voltammetry under standardized conditions.
Abstract
Uric acid (UA), a key purine metabolite associated with metabolic, cardiovascular, and renal disorders, has been extensively investigated using electrochemical sensors. Increasing efforts now focus on translating these technologies into commercial point-of-care (POC) platforms for rapid, decentralized detection. Commercial screen-printed electrodes (SPEs) are widely employed as substrates for such platforms, typically after surface modification with nanomaterials or recognition elements; however, their inherent, unmodified analytical capabilities remain largely uncharacterized. This gap is significant, as the baseline performance of the underlying electrode ultimately constrains or enables the benefits of any subsequent modification. Here, we systematically compare two commercially available SPEs, DRP-110 and DRP-150, for UA detection using differential pulse voltammetry under standardized conditions. Both electrodes exhibited two distinct UA oxidation peaks. The first peak gave linear calibration over 0.02–0.5 µM (R2 = 0.9863 DRP-110, 0.9928 DRP-150), with validated LOD/LOQ of 0.076/0.230 µM (DRP-110) and 0.055/0.166 µM (DRP-150); the second peak extended the linear working range to 1–1000 µM (DRP-110, R2 = 0.9833) and 1–2000 µM (DRP-150, R2 = 0.9928), with LOD/LOQ of 3.37/10.20 µM and 2.23/6.77 µM, respectively, derived from its low-concentration segment. DRP-150 demonstrated superior sensitivity at both peaks. However, both electrodes showed low selectivity toward ascorbic acid. Spike recovery testing in modified artificial urine gave recoveries of 68.0–98.49%, supporting preliminary feasibility in a defined urine-like matrix.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Determination of Uric Acid in Human Urine by Differential Normal Pulse Voltammetry With a Bare Glassy Carbon ElectrodeElectroanalysis · 2026
- Analytical quality by design for electrochemical nanostructured screen-printed sensor development: uric acid detection in saliva for diagnosis and monitoring of chronic diseases.Talanta · 2026
- Redox-Active Polyindole-Modified Electrode for Electrochemical Uric Acid Determination.ACS omega · 2026
- Poly(bromocresol green) modified screen-printed carbon electrode for selective voltametric detection of uric acid.RSC advances · 2026
- A 3D-printed hand-held visible spectrophotometric detector for point-of-care uric acid detection.Spectrochimica acta. Part A, Molecular and biomolecular spectroscopy · 2026
- Enzymatic electrochemiluminescence sensor based on a ternary luminol–H₂ O₂ -confined nanocatalyst system for sensitive determination of uric acidRSC Advances · 2026