Redox-Active Polyindole-Modified Electrode for Electrochemical Uric Acid Determination.
Alam Zeb, Salma Bilal, Bushra Begum, Anwar Ul Haq Ali Shah, Philipp Röse
National Centre of Excellence in Physical Chemistry, University of Peshawar, Peshawar, Khyber Pakhtunkhwa 25120, Pakistan. · Institute of Chemical Sciences, University of Peshawar, Peshawar, Khyber Pakhtunkhwa 25120, Pakistan.
Key points
From the paper's conclusion- These findings identify polyindole as a promising, still underexplored intrinsically conducting polymer for electrochemical biosensing.
Abstract
Uric acid (UA) is an important clinical biomarker whose reliable determination in complex biological media is often hampered by coexisting electroactive interferents. Intrinsically conducting polymers are versatile materials for electrochemical sensing, yet polyindole (PIn) has remained largely unexplored for this purpose. Here, we report a PIn-modified glassy carbon electrode for the nonenzymatic determination of UA in neutral phosphate buffer. PIn was synthesized by chemical oxidative polymerization in the presence of sulfonate dopants and deposited as a thin film on glassy carbon; physicochemical characterization confirmed a doped, conjugated, and nanostructured coating with a high effective surface area. The electrode showed a well-defined UA oxidation signal and, by chronoamperometry, a linear response over 0.01-0.10 mM with a sensitivity of 4.53 μA cm-2 mM-1 and a limit of detection of 4.0 μM. Under the applied conditions, ascorbic acid, dopamine, glucose, and sodium chloride contributed only marginally to the response, and the sensor performed reliably in diluted human serum, with recoveries of 99.7-99.9% and relative standard deviations below 5%. These findings identify polyindole as a promising, still underexplored intrinsically conducting polymer for electrochemical biosensing.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluating Analytical Performance of Commercial Screen-Printed Electrodes for Point-of-Care Uric Acid Sensing: A Comparative StudyApplied Sciences · 2026
- Poly(Allura Red AC)-Modified Glassy Carbon Electrode: Electropolymerization, Characterization, and Application in Uric Acid Detection.ACS omega · 2026
- Poly(bromocresol green) modified screen-printed carbon electrode for selective voltametric detection of uric acid.RSC advances · 2026
- Determination of Uric Acid in Human Urine by Differential Normal Pulse Voltammetry With a Bare Glassy Carbon ElectrodeElectroanalysis · 2026
- AuNPs/reduced graphene oxide biointerface with immobilized uricase for mediator-free electrochemical uric acid biosensing.Analytical and bioanalytical chemistry · 2026
- An Electrochemical Sensor Based on CoFe 2 O 4 /R‐GO Nanocomposite Modified Screen‐Printed Electrode for the Detection of Uric Acid: Mechanistic Insights and Response Surface OptimizationChemistrySelect · 2026