A-228 Multivariable Statistical Analysis Techniques to Identify Subpopulations in Method Comparisons Studies. A Practical Example Comparing with Two Methods for Erythrocytes Sedimentation Rate.
Kathryn Golab, Wazhan Tang, Ryan Montgomer, Tewodros Mola, Dina Radi, Valerio Genta
Key points
From the paper's conclusion- In conclusion, the multivariable regression analysis model technique offered an immediate, efficient and effective method to identify subpopulations in the patient population used for this comparison study.
Abstract
Background Erythrocytes sedimentation rate methods comparisons employing patient specimens may be affected by subpopulations of patient’s specimens characterized by different rates. While these subpopulations may be detected with multiple comparisons employing univariate data analysis techniques, their detection with multivariable data analysis techniques is preferable for efficiency and avoidance of type one error. We present a practical example for the erythrocytes sedimentation rate performed with two methods employing patient’s blood specimens. Methods Two-hundred-forty-seven individual patient’s blood specimens, collected by venipuncture in lavender tubes (Becton-Dickinson), were assayed with miniiSED® (Alcor Scientific, new method, SVBGH, SBLH), iSEDElite® (Alcor scientific, new method, SCPH) and Excyte® (EliTech Group, current method) in parallel and within fifteen minutes in three Sentara Laboratories (SBLH: 59 specimens, SCPH:68 specimens, SVBGH:120 specimens). The data were transferred manually from the instrument printout to Minitab® (Version 22, Minitab Inc.) statistical software. Demographics and the values for HCT and MCV were recorded for each patient. HCT and MCV values were coded as N for normal, L for lower and H for higher, according to their value in relation to the reference range. The data were then analyzed with univariate ordinary least squares (OLS), orthogonal least squares (OrthLS), weighted polynomial least squares (WPLS) statistical techniques, their diagnostics and graphic representations. Results The univariate OLS (Y=7.6+0.8X, Sy/x=17.5) and OrtLS (Y=2.4+0.97X, Sy/x=13) showed similar regression equations. The plots of the standardized residuals by the fitted value showed a fan-shaped increase in variance for increasing RSR values. To investigate whether the increase in variance was due to patient’s subgroups defined by Location, HCT and MCV coded values, the data were analyzed with WPLS model. The ANOVA table of the WPLS showed probabilistically significant differences for subgroups defined by the HCT (F=16.8, P<0.0001), MCV (F=7.0,P=0.024) values and Location (F=3.8, P=0.024). These results indicated that HCT and MCV were more important predictors than Location and were corroborated by forward and backward selection techniques. The plots of the standardized deleted residuals showed a quasi-normal distribution with a few outliers (Standardized deleted residual>|3|) and one influential observation (Hi=0.7). The plot of the miniiSED values by the Excyte values showed four distinct regression lines for LHCT-LMCV, LHCT-NMCV , LHCT-HMCV which fanned away from the NHCT-NMCV regression line. This explained the wide distribution of the observations for increasing ESR values. Conclusion The WPLS model identified the effect of abnormal HCT and MCV values on the ESR rate. This was clearly illustrated by graphic representation of the regression lines for the normal and abnormal combinations of HCT and MCV. Consequently, only the subpopulation characterized by normal HCT and MCV values may have interchangeable ESR reference values. However, additional studies with a larger patient population may be required to define the ESR reference values for the subpopulations with abnormal HCT and MCV. In conclusion, the multivariable regression analysis model technique offered an immediate, efficient and effective method to identify subpopulations in the patient population used for this comparison study. Finally, the availability of univariate and multivariable statistical software such as Minitab was critical.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Analytical Agreement Between the Vision C Automated ESR Analyzer and the Manual Westergren Reference Method: A Prospective Method Comparison Study.Diagnostics (Basel, Switzerland) · 2026
- How to Establish a Reference Interval: A Practical Approach.Clinical chemistry · 2026
- Comparison of Westergren and Automated Erythrocyte Sedimentation Rate Measurements in Dogs and Cats.Animals : an open access journal from MDPI · 2026
- Evaluation of volumetric DBS for lead analysis using ICP-MS/MS: addressing matrix effects and hematocrit variability.Bioanalysis · 2026
- E-norms at Ten: What a Decade Taught the Method About Deriving Reference Values from Patient DatamedRxiv · 2026
- Performance Assessment of Centrifugation-Based Hematocrit Testing in Primary Care Units: Evidence from Seventy-Two Proficiency Testing Datasets over Eight Years.Diagnostics (Basel, Switzerland) · 2026